What molecules have an absorption maximum of 230 nm?

Can I use the BioPhotometer for protein determination in plasma following the Biuret method?

I work with the Bradford protein determination method. During my latest sample measurement on the BioPhotometer the measured absorbance appeared on the display, but a concentration wasn't displayed. What causes this?

How long is the cable you sell for connection of the BioPhotometer to a PC (included in BioPhotometer software package)?

I would like to carry out a colorimetric detection on the BioPhotometer with several standards and using linear regression. Following the measurement, can I query the factor or the ascent of the calibration lines on the BioPhotometer?

The measurement of the optical density of my turbid bacterial culture resulted in a concentration of nil. Why?

How can I calculate as exactly as possible the factor for the photometric determination of concentration of oligonucleotides?

Can I measure the concentration of ring-shaped DNA on the BioPhotometer?

In your "Tips & Tricks" list for the photometric determination of nucleic acid concentrations you say under "Cuvette handling" that for very exact measurements a cuvette error should be determined and allowed for in the measured results. How do I determine the cuvette error?

How can I clean the cuvette shaft of my BioPhotometer?

Which detector is found in the BioPhotometer?

I use the BioPhotometer in the course of my workday at irregular intervals. Is it necessary to switch the device on and off for each measurement or can I leave it on over a period of time?

In the course of my DNA concentration determination I measured a relatively low A260 value (between 0.05 and 0.07) and a relatively high A230 value (corresponding app. to the A260 values). The A260/A230 is <2.0. What do these values say about the quality of my DNA?

Can I measure double-strand RNA on the BioPhotometer?

Our lab works according to GLP (Good Laboratory Practice). What documentation options do I have with respect to quantitative sample measurements on the BioPhotometer?

Can I use the BioPhotometer Data Transfer Software also with the BioPhotometer 6131?

My computer has no serial ports, but only USB connection. Is it possible to work with the BioPhotometer Data Transfer Software nevertheless?

During the start of the BDTS program I get an error message! What can be the reason?

I would like to connect my results to a series of measurement. Is this possible with the new software?

In the area "archive" are a huge amount of measurement values. Can I use search function in the software.

What happened with my results in the area "work". Do I have to export this measurements to the archive by myself or does this process happen automatically?

How can I create a new "archive"?

Is it possible with the new BDTS to create specific user profile?

In how many languages will the new BDTS be offered?

Will the new BDTS software also work Apple Macintosh computer?

What are the system requirements for the BDTS?

At what height does the light beam pass through the cuvette in the BioPhotometer?

For my DNA concentration determination (A260) the BioPhotometer shows a negative value at 230 nm and as a result no ratio A260/A230. What causes the negative A230 value?

Does the OD 600 method take place linearly on the BioPhotometer?

How can I convert an OD 600 value to cell number/ml?

Parallel measurements of my DNA sample showed concentration variations within the individual measurement values. What can cause this?

I have installed your PC online program for the transfer of my measurement data from the BioPhotometer to a PC on my computer. Unfortunately the header lines seem to be missing. What happened?

Which buffer do you recommend using as a measuring medium for the photometric measurement of nucleic acids?

I get fluctuating values for repeated measurements within the Bradford method on the BioPhotometer. I work with a plastic cuvette.

What are the dimensions and weight of the Thermal printer DPU 414?

Which cable is required for connecting the Thermal Printer DPU 414 to the BioPhotometer?

Can I also connect printers other than the Thermal Printer DPU 414 to the BioPhotometer?

Where can I get my Secondary UV-VIS-Filter set recalibrated?

Can I print out data in Excel format over the RS-232 interface of the BioPhotometers?

How often should the BioPhotometer be checked with the Secondary UV-VIS-Filter set?

Can I save several calibration curves per protein determination method in the BioPhotometer (Bradford, Lowry, BCA) ?

What is the smallest filling volume of a cuvette that can be measured on a BioPhotometer?

What is the smallest quantity of DNA that can reasonably be measured with the BioPhotometer?

What is the smallest quantity of RNA that can reasonably be measured with the BioPhotometer?

Can I record standard curves on the BioPhotometer?

Why is a different factor used to calculate the concentration of double-stranded DNA than for calculating the concentration of single-stranded DNA?

How are the filters "Sample 260 nm" and "Sample 280 nm" differentiated from the "Sample A1", "Sample A2" and "Sample A3" filters of the Secondary UV-VIS-Filter set?

Can the measurement values from the BioPhotometer be transferred to a PC?

I would like to perform precision measurement with my BioPhotometer using the "Function" menu. What should I bear in mind here and which solutions are suitable for such measurement?

How long is the VGA printer cable you sell (order no. 0013 610.517) for connection of the BioPhotometer with the serial printer DPU 414?

In your "Tips & Tricks" list for the photometric determination of nucleic acid concentrations you recommend mixing samples in a thermomixer. Can I also use a vortexer for this?

Is the PC online program of your BioPhotometer software package compatible with Windows NT 4.0?

Is the BioPhotometer PC online program compatible with Windows XP?

During my blank measurement on the BioPhotometer only "++++" appears on the display instead of 0.000 A. What happened?

Are batch-specific certificates for the new purity grades available in the internet?

Why are there double labels on the folding boxes, as "PCR clean" + "Sterile" or "PCR clean" + "Protein-free"?

Are Eppendorf products free of RNA?

Why are there new Eppendorf purity grades in 2012?

Have the products changed with the new purity grades?

Which purity grades will be modified from 2012?

Can products of the previous and new modular purity grades (with new seals) be used simultaneously in one application?

What is the UVette made of?

Does the UVette material contain halogens (fluorine, chlorine, bromine, iodine), polystyrene, softeners, phosphorus, nitrogen or metals?

When do you recommend using the 2-mm light path of the UVette and when should the 10-mm light path be used?

Which reagents is the UVette resistant to, and which reagents is it not resistant to?

I would like to use the UVette in a spectrophotometer. How can I determine which adapter I need for my device?

In what temperature range can the UVette be used?

Is the UVette resistant to reagent xyz?

Can I reuse the UVette?

Can I also perform my sample measurement with an optical path length of 2 mm when I use the UVette in combination with an adapter?

My application requires a sterile cuvette. Is the UVette autoclavable?

Do I need to place the UVette in a certain direction in a spectrophotometer?

Should the same UVette be used for the measurement of the blank and of the sample?

What does the term "light path height" mean?

What is important to know when I am working with the UVette with a minimum fill volume of 50 µl?

What are the dimensions of the individual UVette adapters?

What significance does the arrow with the embossed "10" have for the UVette?

What is the maximum filling volume of the UVette?

What is the minimum filling volume of the UVette?

Are the individual UVettes blister-packed in a specific direction?

What is the basic absorbance (self-absorbance) of the UVette?

How can I avoid possible contamination of my highly-sensitive probes as a result of carrying out the previous blank measurement in the same UVette?

Is a transmission curve (self-absorption of the UVette at different wavelengths) available for the UVette?

What kind of applications can be used with method “Assay 490”?

I cannot find the method key “Bradford” for protein determination in the BioPhotometer plus! Is this method still available in the BioPhotometer plus?

How many methods are available in the BioPhotometer plus in contrast to the BioPhotometer?

What are the additional wavelengths that I can use with the new BioPhotometer plus in contrast to the old BioPhotometer?

I am working with dsDNA method. I have seen in the result display of the BioPhotometer plus that the turbidity correction is carried out at 340 nm instead of 320 nm. Why?

What is the meaning of the “Dye 550” and “Dye 650” method at the BioPhotometer plus?

What is the meaning FOI?

What is the recommended FOI value of a sample (e.g., cDNA) for a sufficient signal on the microarray?

Can I change the concentration unit for the fluorescent dye in the “DyeMethods”?

What is the correction factor 260 and 280 in the “Dye 550/650 methods for?

What kind of applications can be used with the method “Assay 340”?

What kind of applications can be used with the method “Assay 405”?

Why can I change the optical pathlength to 0.2 mm in the new Biophotometer plus?

What is the key “absorbance” in the new BioPhotometer plus for?

What kind of wavelengths can be used with the BioPhotometer plus?

How many wavelengths can be used with new BioPhotometer plus?

Can you run kinetics with the new BioPhotometer plus in contrast to the former BioPhotometer?

Can I use the "Warburg-Formula" with the BioPhotometer plus for determining the protein concentration at 280 nm?

Are there any preprogrammed Method in the Eppendorf BioSpectrometer compared to the BioPhotometer plus?

Which cuvettes can be recommended for the Eppendorf BioSpectrometer?

Can the Hellma® TrayCell be used for micro volume measurements in the Eppendorf BioSpectrometer?

Can I use the Eppendorf UVette® in the Eppendorf BioSpectrometer in the same way as I did in the BioPhotometer plus ?

Which cuvettes are recommended for a temperature control in the Eppendorf BioSpectrometer kinetic?

Can the Eppendorf UVette be used for temperature-dependent measurements in the Eppendorf BioSpectrometer kinetic?

What are the differences in the Eppendorf BioSpectrometer kinetic between the “Advanced kinetics” and “Simple kinetics” method?

Can different absorbance spectra be compared directly together in the Eppendorf BioSpectrometer?

Can I compare in the Eppendorf BioSpectrometer my recorded spectrum directly with the example spectra stored in the device?

I have read that in the Eppendorf BioSpectrometer it is possible to export my results to an USB-Stick or PC. In what file format the data are exported?

Can the Eppendorf BioSpectrometer be connected to the internal network and is it possible to print via network printers?

Can methods re-imported to the Eppendorf BioSpectrometer?

Can I use my BioPhotometer filter set for the verification of the BioSpectrometers?

Can the Eppendorf µCuvette also be used to measure samples that contain detergents

My Eppendorf µCuvette has a bit play in the cuvette shaft. Is this normal?

Is it possible to use the Eppendorf µCuvette also for a colorimetric assay , e.g. to determine protein concentrations?

What is the maximum amount of DNA concentration that could be detected with the Eppendorf µCuvette?

What is the maximum amount of protein concentration that could be detected with the Eppendorf µCuvette?

What parameters I need to change in the BioPhotometer plus to measure with the Eppendorf µCuvette?

What parameters I need to change in the BioSpectrometer to measure with the Eppendorf µCuvette?

What do I have to consider for cleaning the Eppendorf µCuvette?

After adding the sample to the Eppendorf µCuvette no liquid column could be formed. What could be the reason for this?

The device shows after the measurement lower concentrations as expected. What could be the reason for this?

Can the Eppendorf µCuvette also be used for spectrophotometers from other manufacturers (not Eppendorf)?

How can I see in the measuring results that no liquid column was formed?

Why I have to use more volume for a protein solution for a measurement as for nucleic acid solution?

I have an old BioPhotometer 6131. Unfortunately I have no possibility to set the 1 mm path length in the parameter section. What can I do?

For what kind of Eppendorf Photometer could the Eppendorf µCuvette be used?

I have measured an absorbance of 1,2 with a cuvette with 10 mm path length. In which range can I expect the absorbance when I measure the same sample in the µCuvette?

Do you recommend anything to prevent contaminations by nucleic acids?

In which direction I have to insert the Eppendorf µCuvette into the cuvette shaft?

Can the Eppendorf µCuvette be cleaned in an ultrasonic bath?

Is it possible to use the Eppendorf µCuvette also in the BioPhotometer D30?

What is the minimum volume for nucleic acid solution that I can use in the Eppendorf µCuvette that I can use in the BioPhotometer D30?

What is the minimum volume for protein solution that I can use in the Eppendorf µCuvette that I can use in the BioPhotometer D30?

Is it possible to carry out a wavelength scan with the BioPhotometer D30?

What wavelength could be applied in the BioPhotometer D30?

What methods are preprogrammed in the BioPhotometer D30?

Is it possible to use the BioPhotometer Data Transfer Software for the BioPhotometer D30?

What can I do with the so called "impurity scan"?

For what methods the "impurity scan" could be applied?

What wavelength range is used for the "impurity scan"?

Is it possible to change the range for the "impurity scan" and can I display the single absorbance values of the "impurity scan"?

Can I use the Hellma® TrayCell with the BioPhotometer D30?

What kind of cuvette path lengths could be applied for the BioPhotometer D30?

Is there any possibility to check the BioPhotometer D30?

In what file-format are the files from the BioPhotometer D30 exported?

Is it possible to select my data for the export?

What kind of SI-units can be applied with the BioPhotometer D30?

In the BioPhotometer D30 certain factors are preprogrammed for the evaluation of specific proteins. Is it possible to apply factors also for other proteins?

What is the meaning of the value "A0,1%"?

How many methods can be stored in the BioPhotometer D30?

How many results can be stored in the BioPhotometer D30?

What molecules have an absorption maximum of 230 nm?Back to top
Carbohydrates, aromatic groups, phenols and peptides have an absorption maximum of 230 nm.
Can I use the BioPhotometer for protein determination in plasma following the Biuret method?Back to top
Because the Biuret method can also be carried out at 562 nm (the absorption behavior of the Biuret reagent is almost identical in the range from 530 to 570 nm), you can measure your Biuret method on the BioPhotometer using the BCA method. Please note that you need only work with a standard for the Biuret method. The measurement of a standard curve is unnecessary.
I work with the Bradford protein determination method. During my latest sample measurement on the BioPhotometer the measured absorbance appeared on the display, but a concentration wasn't displayed. What causes this?Back to top
Please check whether:
- the correct concentration unit was selected (e.g. mg/ml was chosen, but the sample has a concentration in the µg/ml range)
- an entry with decimal places was made for the standard concentrations. With an entry without decimal places, lesser sample concentrations (e.g. 0.024) cannot be recognized.
How long is the cable you sell for connection of the BioPhotometer to a PC (included in BioPhotometer software package)?Back to top
It is 3 m long.
I would like to carry out a colorimetric detection on the BioPhotometer with several standards and using linear regression. Following the measurement, can I query the factor or the ascent of the calibration lines on the BioPhotometer?Back to top
Yes. For each colorimetric calibration saved on the BioPhotometer, all calculated parameters, such as, for example, the ascent of the calibration lines, are automatically filed in the affiliated calibration report. To call up the calibration report press the Function key, select "Calibration report" with the cursor, press the Enter key, select the desired calibration with the cursor, press the Enter key.
The measurement of the optical density of my turbid bacterial culture resulted in a concentration of nil. Why?Back to top
Please check whether the factor 1.000 is found under "Parameter" on the BioPhotometer. It is possible that the factor was accidentally reset to nil when working with the BioPhotometer. In case you continue to have difficulties with measurement, please contact our Application-Hotline: Tel. +49 180 3 66 67 89; email: support@eppendorf.com.
How can I calculate as exactly as possible the factor for the photometric determination of concentration of oligonucleotides?Back to top
In general, the following applies: F = 1: (e x l) where F = factor [(µg/ml)-1 cm-1], e = molar extinction coefficient [M-1 cm-1], l = optical layer thickness of cuvette [cm]. For small molecules like oligonucleotides, for example, the correct extinction coefficient is determined from the base composition. As the concentration of oligonucleotides is commonly reported as mmol/liter, a millimolar extinction coefficient (E) is conventionally used in the Beer-Lambert equation by means of: E = A (15.3) + G (11.9) + C (7.9) + T (9.3) A, G, C and T here stand for the number of corresponding bases in this oligonucleotide, the numbers in brackets for the molar extinction coefficient of any deoxynucleotide at pH 7.0.
Source: Sambrook et. al. Molecular Cloning (2001).Third Edition, A8.21
Can I measure the concentration of ring-shaped DNA on the BioPhotometer?Back to top
Yes. Double-stranded ring-shaped DNA can be measured on the BioPhotometer using the "dsDNA" method, and single-stranded ring-shaped DNA with the "ssDNA" method.
In your "Tips & Tricks" list for the photometric determination of nucleic acid concentrations you say under "Cuvette handling" that for very exact measurements a cuvette error should be determined and allowed for in the measured results. How do I determine the cuvette error?Back to top
You fill a cuvette with water and measure it as blank. Now fill another cuvette with water as well and measure this cuvette as a sample. The difference in absorption between the two cuvettes can now be identified as positive or negative absorption at the individual wavelengths. The most reliable method is to repeat the sample measurement up to three times and then to calculate the cuvette error as the mean value of the absorption differences measured.
How can I clean the cuvette shaft of my BioPhotometer?Back to top
Dust in the cuvette shaft can be carefully removed with a cotton tip. If the cuvette shaft is heavily soiled, for example with dried-on buffer residues, the cotton tip should be carefully soaked in 40% methanol beforehand.
Which detector is found in the BioPhotometer?Back to top
Silicium photo diodes are used as detectors in the BioPhotometer.
I use the BioPhotometer in the course of my workday at irregular intervals. Is it necessary to switch the device on and off for each measurement or can I leave it on over a period of time?Back to top
The life span of the xenon lamp is influenced by neither repeated switching on and off nor by leaving it on in standby over a period of several hours. You can therefore decide for yourself. The xenon lamp is also only active during the measurement. It then switches over into standby mode.
In the course of my DNA concentration determination I measured a relatively low A260 value (between 0.05 and 0.07) and a relatively high A230 value (corresponding app. to the A260 values). The A260/A230 is <2.0. What do these values say about the quality of my DNA?Back to top
Please note that you should measure at 260 nm for a minimum absorbance of 0.1 in order to obtain exact results. This value is independent of the measuring device and is based upon the disturbing influence of impurities and particles on the measurement result, which is especially high at A < 0.1. An A260/A230 < 2.0 can indicate impurity of the DNA sample. This is because peptides, aromatic groupes, phenols and carbohydrates absorb at 230 nm. With contaminated DNA samples you can try to clean the sample through centrifugation (at least. 5 min). Otherwise the sample must be purified again according to the protocol specifications.
Can I measure double-strand RNA on the BioPhotometer?Back to top
Yes, if you are aware of a factor for double-strand RNA. You should enter this under the RNA method before the measurement under "Parameters". If you don't have a factor, you can carry out a calibration with a familiar dsRNA quantity and then use this to measure the unknown sample.
Our lab works according to GLP (Good Laboratory Practice). What documentation options do I have with respect to quantitative sample measurements on the BioPhotometer?Back to top
For purposes of data documentation, either a printer or a PC can be connected to each BioPhotometer. It is also possible to have the photometric systematic (trueness) and photometric random error (precision) of your device inspected by one of our service partners. For purposes of testing directly, we offer the user a Secondary UV-VIS-Filter.
Can I use the BioPhotometer Data Transfer Software also with the BioPhotometer 6131?Back to top
Yes, the software is valid for the BioPhotometer plus as well as for the BioPhotometer 6131!
My computer has no serial ports, but only USB connection. Is it possible to work with the BioPhotometer Data Transfer Software nevertheless?Back to top
Yes, this is possible. You need an RS232/USB connection cable. This cable is worldwide available by the company VScom.
During the start of the BDTS program I get an error message! What can be the reason?Back to top
Please check, if all systems settings as stated in the manual in chapter 2.3 are correct.
I would like to connect my results to a series of measurement. Is this possible with the new software?Back to top
Yes, this is possible. Click in the "work" area on "ID". Here you can connect you results to a specific series of measurement!
In the area "archive" are a huge amount of measurement values. Can I use search function in the software.Back to top
Yes, this is possible. Therefore you can use in the area "archive" the filter function.
What happened with my results in the area "work". Do I have to export this measurements to the archive by myself or does this process happen automatically?Back to top
No, this data will be exported automatically.
How can I create a new "archive"?Back to top
Click top menu bar on "new archive file"!
Is it possible with the new BDTS to create specific user profile?Back to top
Yes, this is possible! Just click on top menu bar on "file" and "settings". In the right window you can apply a new user profile with a new password. Please note that only a new profile can apply if you are login as an administrator.
In how many languages will the new BDTS be offered?Back to top
The BDTS will only be offered in English language.
Will the new BDTS software also work Apple Macintosh computer?Back to top
No, the BDTS will only be available for Windows.
What are the system requirements for the BDTS?Back to top
Windows XP or Windows Vista.
At what height does the light beam pass through the cuvette in the BioPhotometer?Back to top
At 8.5 mm.
For my DNA concentration determination (A260) the BioPhotometer shows a negative value at 230 nm and as a result no ratio A260/A230. What causes the negative A230 value?Back to top
Negative A230 values are usually caused by interference components in inadequately concentrated DNA solutions. The negative value should be rectified when you use a lesser sample dilution for the next measurement. Please note that you should measure at 260 nm for a minimum absorbance of 0.1 in order to obtain exact results. This value is independent of the measuring device and is based upon the disturbing influence of impurities and particles on the measurement result, which is especially high at A < 0.1.
Does the OD 600 method take place linearly on the BioPhotometer?Back to top
This measuring method is principally non-linear, when one observes the absorption curve over wide OD ranges and dilutions. The values are dependent on both the organism (size, shape) and the device (light path geometry, ratio of measured stray light radiation). A correct procedure includes the generation of a calibration graph for each organism and device by determining the actual number of cells under a microscope. When the curve shape is known it is then possible to measure in ranges that are almost linear.
How can I convert an OD 600 value to cell number/ml?Back to top
Conversion is possible through calibration graphs or counting. With measurements of Escherichia coli, for example, it is possible to estimate roughly that an OD 600 between 0.5 and 1.0 corresponds approximately to a cell number between 1 x 108 and 1 x 109.
Parallel measurements of my DNA sample showed concentration variations within the individual measurement values. What can cause this?Back to top
Please check whether
1.The absorbance of your DNA sample is greater than/equal to 0.1 (please note that regardless of the device used, absorption measurements of nucleic acids only deliver reliably reproducible results at absorbances of greater than 0.1. This is based upon the disturbing influence of impurities and particles on the measurement result, which is especially high at A < 0.1.)
2. Your sample was sufficiently mixed (only careful mixing of the sample can guarantee consistent measurement values)
3. There is an impurity in your sample. An indicator of a possible impurity is the relative ratio of A260/A280 or A260/A230, which with pure DNA should amount to app. 2.0 or 2.5 at pH 7-8.5. Carbohydrates, aromatic groups, phenols and peptides have an absorption maximum of 230 nm, proteins of 280 nm.
Additional literature: "Tipps and Tricks for photometric quantification of nucleic acids".
I have installed your PC online program for the transfer of my measurement data from the BioPhotometer to a PC on my computer. Unfortunately the header lines seem to be missing. What happened?Back to top
Please switch the BioPhotometer off and then on again following the installation of the program. The header line will then appear.
Which buffer do you recommend using as a measuring medium for the photometric measurement of nucleic acids?Back to top
Wilfinger et. al* generally recommend using 1 - 3 mM Na2HPO4, pH 8.0 - 8.5 as the measuring medium for nucleic acids. For DNA measurements, our experience shows that 10 mM Tris-HCl, pH 8.0 or a TE buffer, pH 8.0 are also suitable. *Wilfinger W.W., Mackey K. and Chomczynski P. 1997. Effect of pH and ionic strength on the spectrophotometric assessment of nucleic acid purity. BioTechniques 22: 474-481.
I get fluctuating values for repeated measurements within the Bradford method on the BioPhotometer. I work with a plastic cuvette.Back to top
The Bradford reagent is an aggressive substance that attacks the plastic of plastic cuvettes when the exposure time is too long. This reagent should therefore not remain in a plastic cuvette for more than 5 minutes. Fluctuating values could be caused by the fact that the Bradford reagent has already been too long in the cuvette.
Tip: Use glass cuvettes for the Bradford method.
What are the dimensions and weight of the Thermal printer DPU 414?Back to top
Dimensions (WxHxD) = 160 mm x 70 mm x 170 mm; Weight app. 400 g.
Which cable is required for connecting the Thermal Printer DPU 414 to the BioPhotometer?Back to top
The Thermal Printer is delivered equipped with an appropriate cable. Replacement cables are available from us upon request.
Can I also connect printers other than the Thermal Printer DPU 414 to the BioPhotometer?Back to top
From software version 1.20 on you can also connect other serial and parallel printers (a converter cable is required) besides the Thermal Printer DPU 414 to the BioPhotometer. If you use the BioPhotometer with software version 1.0, you can only connect the Thermal Printer DPU 414.
Where can I get my Secondary UV-VIS-Filter set recalibrated?Back to top
Please send your Secondary UV-VIS-Filter set to your local Eppendorf partner.
Can I print out data in Excel format over the RS-232 interface of the BioPhotometers?Back to top
From software version 1.20 on you can transfer measuring values from the BioPhotometer to a PC (respectively, in an Excel table) with the help of the BioPhotometer software package and then process them further or print them out.
How often should the BioPhotometer be checked with the Secondary UV-VIS-Filter set?Back to top
We recommend checking the photometric systematic error (photometric trueness) and the wavelength systematic error (wavelength trueness) once a year using the Secondary UV-VIS-Filter set.
Can I save several calibration curves per protein determination method in the BioPhotometer (Bradford, Lowry, BCA) ?Back to top
You can save one standard and one micro method for each protein determination method. This means that you can store two calibration curves for each protein determination method.
What is the smallest filling volume of a cuvette that can be measured on a BioPhotometer?Back to top
That depends upon the cuvette chosen. We recommend using the UVette with a minimum volume of 50 µl.
What is the smallest quantity of DNA that can reasonably be measured with the BioPhotometer?Back to top
The BioPhotometer should not be used to measure an absorbance of less than 0.05 at 260 nm. This OD value corresponds to a dsDNA concentration of app. 2.5 µg/ml (= app. 125 ng dsDNA in a 50 µl preparation). Please note that you should measure at 260 nm for a minimum absorbance of 0.1 in order to obtain exact results. This value is independent of the measuring device and is based upon the disturbing influence of impurities and particles on the measurement result, which is especially high at A < 0.1.
What is the smallest quantity of RNA that can reasonably be measured with the BioPhotometer?Back to top
The BioPhotometer should not be used to measure an absorbance of less than 0.05 at 260 nm. This OD value corresponds to a RNA concentration of app. 2 µg/ml (= app. 100 ng RNA in 50 µl preparation). Please note that you should measure at 260 nm for a minimum absorbance of 0.1 in order to obtain exact results. This value is independent of the measuring device and is based upon the disturbing influence of impurities and particles on the measurement result, which is especially high at A < 0.1.
Can I record standard curves on the BioPhotometer?Back to top
Yes. This is possible with the Bradford/Bradford micro, Lowry/Lowry micro and BCA/BCA micro methods.
Why is a different factor used to calculate the concentration of double-stranded DNA than for calculating the concentration of single-stranded DNA?Back to top
Factor F is calculated from the molar absorbance coefficient of a molecule in solution and the optical path length of the cuvette used (see Beer-Lambert law). The absorption behavior of nucleic acids is influenced by the aromatic rings of the bases. Here the molar absorbance coefficient of a double-stranded nucleic acid molecule with which the bases are in close contact with each other, is smaller than of a single-stranded nucleic acid molecule. It thus follows that single-stranded nucleic acid molecules have a higher rate of absorption than double-stranded molecules (Sambrook et al. 2001. Molecular Cloning 3rd edition, A8.20). With an optical path length of 10 mm and under neutral to slightly alkaline measuring conditions an optical density of 1 at 260 nm thus corresponds to approx. 50 µg/ml of double-stranded DNA, 37 µg/ml of single-stranded DNA, 40 µg/ml of RNA and appr. 30 µg/ml of oligonucleotides.
How are the filters "Sample 260 nm" and "Sample 280 nm" differentiated from the "Sample A1", "Sample A2" and "Sample A3" filters of the Secondary UV-VIS-Filter set?Back to top
The 260 nm and 280 nm filter samples are for the testing of the wavelength accuracy, the A1, A2 and A3 filter samples for testing of the photometric accuracy (systematic and random error) of all wavelengths. The values measured with these filters must lie within the given limits. The limiting values are contained in a table found on the inside of the lid of the filter box.
Can the measurement values from the BioPhotometer be transferred to a PC?Back to top
As of software version 1.20 you can easily and comfortably transfer measurement data into a table calculation program such as Excel using our BioPhotometer software package. The PC online program also enables the use of the familiar printout design from the thermoprinter with a PC printer.
I would like to perform precision measurement with my BioPhotometer using the "Function" menu. What should I bear in mind here and which solutions are suitable for such measurement?Back to top
Precision measurement should be carried out using a semimicro-quartz glass cuvette (> 350 µl). Clear and clean solutions are suitable as the measuring medium, for example potassium dichromate (zero adjustment to water) or water (zero adjustment to air). The coefficient of variation (CV) should be smaller/equal to 1% at 1 A in line with the technical data for the BioPhotometer. Please bear in mind that the Secondary UV-VIS filter should be used for exact precision measurement and determination of both the photometric and wavelength accuracy.
How long is the VGA printer cable you sell (order no. 0013 610.517) for connection of the BioPhotometer with the serial printer DPU 414?Back to top
It is 2 m long.
In your "Tips & Tricks" list for the photometric determination of nucleic acid concentrations you recommend mixing samples in a thermomixer. Can I also use a vortexer for this?Back to top
You can of course vortex and centrifuge your sample and then transfer it to a cuvette. Please make sure that the sample is transferred to the cuvette air bubble-free .
Is the PC online program of your BioPhotometer software package compatible with Windows NT 4.0?Back to top
Yes.
Is the BioPhotometer PC online program compatible with Windows XP?Back to top
Yes.
During my blank measurement on the BioPhotometer only "++++" appears on the display instead of 0.000 A. What happened?Back to top
This is caused by the measurement outside of the upper measuring range (> 3.0 A).
Please check
1. Whether your cuvette is pressed firmly onto the base of the cuvette shaft. The error message "++++" indicates that the cuvette base is blocking the light beam.
2. Whether your cuvette has an optical window at 8.5 mm.
3. Whether you have sufficient blank medium in the cuvette.
Are batch-specific certificates for the new purity grades available in the internet?Back to top
Yes, the batch-specific purity grades can be downloaded at www.eppendorf.com/certificates.
Why are there double labels on the folding boxes, as "PCR clean" + "Sterile" or "PCR clean" + "Protein-free"?Back to top
Due to the modularization of purity grades, some products are labeled with more than one purity grade symbol. For example, the epDualfilter T.I.P.S. has the double seal "PCR clean "+ "Sterile" and the UVette the double seal "PCR clean" + "Protein-free" on the packing. The certification remains unaffected by the modularization.
Are Eppendorf products free of RNA?Back to top
All products certified with "PCR clean" are tested free of human DNA, DNase, RNase and PCR inhibitors. The following applies from our point of view: Where no human DNA is detected, there can also be no RNA, which is considerably more unstable.
Why are there new Eppendorf purity grades in 2012?Back to top
The previous purity grades were developed due to customer requirements and new applications, in the course of the years. To meet the customer requirements for single applications even better, Eppendorf has extended the certification of the purity grades "Sterile" (now, also certified as free of pyrogenes) and "Biopur" (now, also certified as free of DNase and PCR inhibitors). In addition, the previous purity grades have been transferred to a new modular system that provides the same variety and flexibility and yet is readily understandable.
Have the products changed with the new purity grades?Back to top
No, they are the same products as before, i.e. all specifications as dimensions remain the same. Only the packing has changed, due to the new seals. In addition, the certification of "Sterile" and "Biopur" has been extended but the entire production process was maintained.
Which purity grades will be modified from 2012?Back to top
For all purity grades, there will be new packing seals. In addition, there will be a further certification for the purity grades "Sterile" and "Biopur". "Sterile" will certify sterility and also the freedom of pyrogenes. The purity grade "Biopur" also includes the freedom of DNase as well as of PCR inhibitors.
Can products of the previous and new modular purity grades (with new seals) be used simultaneously in one application?Back to top
Yes, if no batch attachment is required for the application, products of the old and new modular purity grades can be used, as their specifications are the same.
What is the UVette made of?Back to top
The UVette is made of a UV-transparent plastic.
Does the UVette material contain halogens (fluorine, chlorine, bromine, iodine), polystyrene, softeners, phosphorus, nitrogen or metals?Back to top
No.
When do you recommend using the 2-mm light path of the UVette and when should the 10-mm light path be used?Back to top
In general, the absorbance should be within the linear measuring range of the spectrophotometer (measuring range of Biophotometer: A = 0–3.0). For samples with a normal or a low concentration, the 10-mm light path is more suitable. For high-concentration samples (with A > 2.5), we recommend using the 2-mm light path. (An absorbance > 2.5 correlates to > 125 µg/ml dsDNA and > 100 µg/ml RNA respectively.)
Which reagents is the UVette resistant to, and which reagents is it not resistant to?Back to top
The UVette is resistant to:
- liquid solutions in the pH 0-14 range
- pure acetic acid
- alcohols
- acetone, butanone
- polar organic solvents.
It is not resistant to:
- hydrocarbons
- chlorinated hydrocarbons
- aromatics such as benzene, toluene, xylene.

Please find attached a file for further information.
NameTypeSizeLink
material_stability_uvette.pdfAdobe Acrobat®434 KBDownload
I would like to use the UVette in a spectrophotometer. How can I determine which adapter I need for my device?Back to top
The UVette can in principle be used in devices with which good results have already been attained using ultra micro cuvettes. To select the correct adapter, you need to know the light path height of your photometer. A summary of "UVette adapters" lists the relevant UVette adapters for numerous devices. If your device is not on the list, please ask the device manufacturer for the light path height of your device or contact our Application Hotline (Tel. +49 1803 666 789; support@eppendorf.com).
In what temperature range can the UVette be used?Back to top
The material characteristics of the UVette are identical within a temperature range between room temperature and 40 °C. Measurement of samples between 80 -100 °C is not possible.
Is the UVette resistant to reagent xyz?Back to top
Please take a look into our brochure „Material resistance of the Uvette“, if your reagent is listed. You will find this brochure on our homepage. If your reagent is not mentioned, please ask our application-hotline for assistance (support@eppendorf.com)
Can I reuse the UVette?Back to top
We recommend that the UVette be discarded following the sample measurement in order to guarantee contamination-free analysis. Cleaning processes and/or multiple use of an UVette may damage the quality of the optical surfaces, particularly in the sensitive UV-range.
Can I also perform my sample measurement with an optical path length of 2 mm when I use the UVette in combination with an adapter?Back to top
Yes.
My application requires a sterile cuvette. Is the UVette autoclavable?Back to top
No. This is because autoclaving can result in distortion of or changes to the optical characteristics of the UVette material. However, we deliver the UVette already free of DNA, RNase and protein.
Do I need to place the UVette in a certain direction in a spectrophotometer?Back to top
No. However, as with quartz glass cuvettes, it is important to make certain that the UVette is placed in the same direction in the cuvette shaft for both the blank and the sample measurement.
Should the same UVette be used for the measurement of the blank and of the sample?Back to top
Because the self-absorbance values of the UVettes vary somewhat from one another, ideally, the same UVette should be used for both measurements.
What does the term "light path height" mean?Back to top
The light path height of a photometer informs you at what height, in relation to the bottom of a cuvette, the light beam of the photometer occurs.
What is important to know when I am working with the UVette with a minimum fill volume of 50 µl?Back to top
Please make sure that there are no air bubbles in the solution and that there is no residual liquid on the inner wall of the UVette. Tip: pipette the 50 µl directly into the lower part of the UVette.
What are the dimensions of the individual UVette adapters?Back to top
All UVette adapters have dimensions of 12.5 x 12.5 mm. These correspond to the standard dimensions for cuvettes with 1 cm optical path length.
What significance does the arrow with the embossed "10" have for the UVette?Back to top
The arrow serves as an orientation aid in two ways. First, it clearly shows the 10 mm light beam of the UVette. It also makes it easier to note the selected direction of the UVette in the cuvette shaft for the blank measurement.
What is the maximum filling volume of the UVette?Back to top
Up to 2,000 µl may be pipetted into the UVette.
What is the minimum filling volume of the UVette?Back to top
The minimum filling volume of the UVette is 50 µl.
Are the individual UVettes blister-packed in a specific direction?Back to top
Yes. Each UVette is packaged in such a way that the cuvette opening points toward the tear-strip of the blister-pack. In this way you can be sure not to touch the optical window of the UVette when tearing open the blister-pack.
What is the basic absorbance (self-absorbance) of the UVette?Back to top
The self-absorbance of the UVette is < 0.5; measured with water. The transmission curve is found in the operating manual for the UVette.
How can I avoid possible contamination of my highly-sensitive probes as a result of carrying out the previous blank measurement in the same UVette?Back to top
You can do this by first measuring the sample, then rinsing the UVette thoroughly with buffer before measuring the blank. You must then subtract the blank value from your sample result.
Is a transmission curve (self-absorption of the UVette at different wavelengths) available for the UVette?Back to top
Yes, this can be found in the pack insert of the UVette.
What kind of applications can be used with method “Assay 490”?Back to top
With the Assay 490 method cytotoxicity assays can be evaluated with the BioPhotometer plus via endpoint detection. This assay is applied for testing the vitality of living cells. The assay is based on a lactate dehydrogenase reaction combined with a specific catalyst whereas the endproduct “formazan” can be detected at 490 nm in the BioPhotometer plus. Like in the method “Assay 340” the evaluation of the results could be done via a factor,a standard or a standard curve.
I cannot find the method key “Bradford” for protein determination in the BioPhotometer plus! Is this method still available in the BioPhotometer plus?Back to top
All methods of the BioPhotometer are still maintained. Furthermore in the BioPhotometer plus more methods are available compared to the BioPhotometer. Therefore a reorganization of the keypad was necessary: All related methods are summarized in one group of methods. For instance, in the method group “Protein” you will find all methods for protein determination like the colorimetric assays Bradford, Lowry, BCA as well as protein direct measurements.
How many methods are available in the BioPhotometer plus in contrast to the BioPhotometer?Back to top
With the BioPhotometer plus 32 methods (23 programmed and 9 freely programmable methods) can be used. In the precursor BioPhotometer only 9 methods were available.
What are the additional wavelengths that I can use with the new BioPhotometer plus in contrast to the old BioPhotometer?Back to top
With the new BioPhotometer plus additional wavelengths at 340 nm, 405 nm, 490 nm, 550 nm and 650 nm can be applied. Two of these wavelengths have been changed: 340 nm instead of 320 nm, 550 nm instead of 562 nm. These changes do not effect the performance of any method.
I am working with dsDNA method. I have seen in the result display of the BioPhotometer plus that the turbidity correction is carried out at 340 nm instead of 320 nm. Why?Back to top
Yes, this is correct. Due to this change more applications are possible with the BioPhotometer plus, e.g., endpoint measurement of NADH based enzyme assays. The turbidity correction at DNA, RNA and protein measurements can also be carried out at 340 nm with the same sensitivity.
What is the meaning of the “Dye 550” and “Dye 650” method at the BioPhotometer plus?Back to top
These methods are needed for the detection of fluorescent dye labeled biomolecules, like DNA, RNA or protein. With this application it is possible to check the quality of probes that should be used for microarray experiments. The quality of the probe will be indicated by the FOI (Frequency of incorporation) and the concentration of the incorporated dye.
What is the meaning FOI?Back to top
FOI means “Frequency of incorporation”, indicating the incorporation rate of dye–molecules into a biomolecule (e.g., molecules dye/kb DNA). This function will be available in the methods Dye550 and Dye650, respectively.
What is the recommended FOI value of a sample (e.g., cDNA) for a sufficient signal on the microarray?Back to top
This depends on the fluorescent dye. For cyanine dyes (e.g., Cy3) a FOI value of 15 but not more than 45 molecules fluorescent dyes / 1000 nucleotides is recommended.
Can I change the concentration unit for the fluorescent dye in the “DyeMethods”?Back to top
No. The unit for the dye concentration is defined permanently with "pmol/μl" and can therefore not be changed in the parameters.
What is the correction factor 260 and 280 in the “Dye 550/650 methods for?Back to top
Many common fluorescent dyes, e.g., Cy3 and Cy5, show also little absorbance at 260 and 280 nm. This may lead to minute deviations when measuring the dye labeled DNA, RNA or protein in the BioPhotometer plus. This deviation can be corrected by a correction factor. The correction factors for Cy3, Cy5, ALEXA 555/647 are already programmed. For correction factors for other fluorescent dyes please contact the manufacturer directly.
What kind of applications can be used with the method “Assay 340”?Back to top
This method is used for the evaluation of NADH based enzyme reactions. The enzymatic consumption will be analyzed via endpoint detection. NADH can be detected with the BioPhotometer plus at 340 nm. The evaluation of the results could be done via a factor, a standard or a standard curve.
What kind of applications can be used with the method “Assay 405”?Back to top
This method is for endpoint detection of enzymatic consumption of artificial substrates. These artificial substrates serve as indicators for the activity of specific enzymes, e.g., beta-galactosidase assay or alkaline phosphatase assay. The products of the enzymatical degradation are generally para- or orthonitrophenol (pNP oder oNP). Both chemicals can be detected in the BioPhotometer plus at 405 nm. Like in the method “Assay 340” the evaluation of the results could be done via a factor, a standard or a standard curve.
Why can I change the optical pathlength to 0.2 mm in the new Biophotometer plus?Back to top
With the optical pathlength of 0.2 mm microliter cell like the Hellma ®TrayCell (0.2 mm lid) can be used directly in the BioPhotometer plus without using any “virtual” dilution factors.
What is the key “absorbance” in the new BioPhotometer plus for?Back to top
With the absorbance key it is possible to measure the absorbance of a sample directly without using any calculation parameter.
What kind of wavelengths can be used with the BioPhotometer plus?Back to top
With the BioPhotometer plus the wavelengths 230 nm, 260 nm, 280 nm, 340 nm, 405 nm, 490 nm, 550 nm, 595 nm and 650 nm are available. With the “absorbance” key, all wavelengths can be used for direct measurement without using any calculation factors. The data can be used for further processing.
How many wavelengths can be used with new BioPhotometer plus?Back to top
9 wavelengths can be used with the BioPhotometer plus.
Can you run kinetics with the new BioPhotometer plus in contrast to the former BioPhotometer?Back to top
No, it is not possible to run kinetics in the BioPhotometer plus. However, due to the availability of the additional wavelengths many enzymatic assays can be carried out via endpoint detection.
Can I use the "Warburg-Formula" with the BioPhotometer plus for determining the protein concentration at 280 nm?Back to top
No, this is unfornutately not possible, but in the result display the absorbance at 280 nm and 260 nm is shown. So it is possible to calculate the protein concetration via the "Warburg-Formula" manually: c(Protein) = 1,55 x E(280) - 0,757 x E(260). Please note that photometric measurement of the protein concentrion in the UV-range only suitable for pure and homogeneous protein solutions.
Are there any preprogrammed Method in the Eppendorf BioSpectrometer compared to the BioPhotometer plus?Back to top
Yes, in the method group “Routine” are all methods that also can be found in the BioPhotometer plus. These methods are the detection of nucleic acids, proteins, fluorescent dye labeled biomolecules plus OD600 measurements.
Which cuvettes can be recommended for the Eppendorf BioSpectrometer?Back to top
All cuvettes with standard dimensions can be used (12.5 * 12.5 mm). It is important that at height of 8.5 mm the optical window is transparent for the light beam)? Please check therefore also the specification in the manual (chapter 5.2).
Can the Hellma® TrayCell be used for micro volume measurements in the Eppendorf BioSpectrometer?Back to top
"Yes, this is possible. To consider the light path for the calculation of the sample concentration, this value has to be entered in the area “check parameters”. Please note, that for the use of the TrayCell a sufficient sample concentration has to be applied: For the available sample concentration we recommend the absorbance ranges (e.g. dsDNA): 0.1 mm = 500 – 5000 ng/µL 0:2 mm = 250 – 2500 ng/µL 1 mm = 50 – 500 ng/µL 2 mm = 10 – 100 ng/µL These concentration ranges for dsDNA correspond to an absorbance of 0.1 to 1 A. "
Can I use the Eppendorf UVette® in the Eppendorf BioSpectrometer in the same way as I did in the BioPhotometer plus ?Back to top
Yes, this is easily possible. However, for absorbance measurement below 220 nm you have to consider the self absorbance of the UVette. For this reason we recommend to carry out measurements in the UVette only above 220 nm.
Which cuvettes are recommended for a temperature control in the Eppendorf BioSpectrometer kinetic?Back to top
Basically every cuvette with standard dimensions can be used for the BioSpectrometer kinetic (s. chapter 5.2 in the manual). For optimal temperature control it is important that the cuvette has direct contact with the integrated peltier-element. Most adequate cuvettes are made of glass or quartz glass. In chapter 5.3.4 of the manual an overview is shown regarding pre-incubation times for different cuvette shapes.
Can the Eppendorf UVette be used for temperature-dependent measurements in the Eppendorf BioSpectrometer kinetic?Back to top
Because of the shape of UVette the temperature transfer is here limited we recommend conventional cells made of glass or quartz glass!
What are the differences in the Eppendorf BioSpectrometer kinetic between the “Advanced kinetics” and “Simple kinetics” method?Back to top
In the method, "Advanced kinetics”, the measurement of standards can be programmed. This can be used for the measurement of unknown substrate concentrations via enzyme kinetics. Moreover it is possible if a drift of the reagent solution is expected, a “reagent blank” can be measurement. The result of this measurement is than subtracted from the sample measurement. These measurements are in the method "Simple Kinetics" not possible.
Can different absorbance spectra be compared directly together in the Eppendorf BioSpectrometer?Back to top
No, this is due to patent reasons not possible. But you can export the graph data as an Excel file and then reprocess here accordingly.
Can I compare in the Eppendorf BioSpectrometer my recorded spectrum directly with the example spectra stored in the device?Back to top
No, also this function is due to patent reasons not possible.
I have read that in the Eppendorf BioSpectrometer it is possible to export my results to an USB-Stick or PC. In what file format the data are exported?Back to top
The files will be exported as an Excel file.
Can the Eppendorf BioSpectrometer be connected to the internal network and is it possible to print via network printers?Back to top
No, this is unfortunately not possible. But it is possible to export the data to an USB-Stick. The data can then be printed from the PC over a local or network printer.
Can methods re-imported to the Eppendorf BioSpectrometer?Back to top
A re-import of methods is unfortunately not possible.
Can I use my BioPhotometer filter set for the verification of the BioSpectrometers?Back to top
For the BioSpectrometer an additional filter set will be available.
Can the Eppendorf µCuvette also be used to measure samples that contain detergentsBack to top
No, we cannot recommend to measure samples that contain detergents
My Eppendorf µCuvette has a bit play in the cuvette shaft. Is this normal?Back to top
It depends, of course, on the usage of the Eppendorf µCuvette and the cuvettes shaft but a little bit play is actually normal. In any case the cuvette is automatically shifted to the correct position for the measurement.
Is it possible to use the Eppendorf µCuvette also for a colorimetric assay , e.g. to determine protein concentrations?Back to top
In principle this would be possible. But it has to be kept in mind that the Bradford assay and all other colorimetric assays optimized for cuvettes with a 10 mm optical path length. Since a 10 times less optical path length is used in the Eppendorf µCuvette, the expected absorbance with a colorimetric assay will be close or under the detection limit of the Photometer. This may lead to inaccuracies by creating the standard curve.
What is the maximum amount of DNA concentration that could be detected with the Eppendorf µCuvette?Back to top
Theoretically it is possible to measure up to 1500 µg/mL double-stranded DNA. This would be similar to an absorbance of 3, which is the maximum absorbance that can be measured in the Eppendorf BioPhotometer and BioSpectrometer. For optimal measuring results we recommend to use concentrations up 1000 µg/mL dsDNA (A=2).
What is the maximum amount of protein concentration that could be detected with the Eppendorf µCuvette?Back to top
This depends actually on the protein itself. Theoretically it is possible to measure up to 45 mg/mL BSA. This would be similar to an absorbance of 3, which is the maximum absorbance that can be measured in the Eppendorf BioPhotometer and BioSpectrometer. For optimal measuring results we recommend to use concentrations up 30 mg/mL BSA (A=2).
What parameters I need to change in the BioPhotometer plus to measure with the Eppendorf µCuvette?Back to top
For correct results the settings for the cuvette path length has to be changed to 1 mm in the parameter section: Select first the method family (e.g. DNA). Afterward go with the “cursor” key to your desired method. Press the “parameter” key move the cursor to the cuvette path length of 1 mm. For activating the new path length press the “enter” key. To leave the parameter section press “parameter” again.
What parameters I need to change in the BioSpectrometer to measure with the Eppendorf µCuvette?Back to top
For correct results the settings for the cuvette path length has to be changed to 1 mm. In the BioSpectrometer you may change this setting at the beginning of the method easily in the "check parameters" section. Change here settings for the “cuvette” path length to 1 mm.
What do I have to consider for cleaning the Eppendorf µCuvette?Back to top
Always use for cleaning lint-free tissue papers. Do not use any chemicals that are not recommended in the operation manual. Particularly substances that tend to corrosion such as acids or bases should be avoided.
After adding the sample to the Eppendorf µCuvette no liquid column could be formed. What could be the reason for this?Back to top
One reason can be, that the volume of the used sample was simply too low. Another possibility is that after purification of DNA samples it may appear that there is still high amount of protein inside the sample. This may lead to a reduced surfaced tension of the liquid. This problem could be solved by slightly increasing the measuring volume. If solutions are containing detergent, it may appear that liquid column cannot be formed at all. In this case the sample should be measured in a cuvette (e.g. Eppendorf UVette).
The device shows after the measurement lower concentrations as expected. What could be the reason for this?Back to top
It may be that not the right parameters are adjusted. Please check in the device if the programmed optical path length corresponds with the light path of the Eppendorf µCuvette.
Can the Eppendorf µCuvette also be used for spectrophotometers from other manufacturers (not Eppendorf)?Back to top
Unfortunately we cannot recommend this. The Eppendorf µCuvette is specialized and optimized for the usage in Eppendorf Photometer.
How can I see in the measuring results that no liquid column was formed?Back to top
Please check in the measuring results, e.g. if you want to determine the DNA concentration in your sample, if the purity ratios (260/280 or 260/230) are unusually high or low. Also a high background signal indicated by a high absorbance at 320 or 340 nm (A>0,1), can be a sign for a non-formed liquid-column. Please check also the scan result when using a BioSpectrometer for your measurement. If during DNA-measurements a typical absorbance curve is not displayed, the liquid column should be checked, too.
Why I have to use more volume for a protein solution for a measurement as for nucleic acid solution?Back to top
Proteins can reduce the surface tension in aqueous solutions, so that liquid column cannot correctly be formed. To increase the volume may help to form the liquid column completely.
I have an old BioPhotometer 6131. Unfortunately I have no possibility to set the 1 mm path length in the parameter section. What can I do?Back to top
Alternatively you may program also a dilution. Programming a 1:10 dilution before the measurement will lead to the same result.
For what kind of Eppendorf Photometer could the Eppendorf µCuvette be used?Back to top
The Eppendorf µCuvette can be used for all Eppendorf BioSpectrometer and BioPhotometer.
I have measured an absorbance of 1,2 with a cuvette with 10 mm path length. In which range can I expect the absorbance when I measure the same sample in the µCuvette?Back to top
Since a 10 times smaller optical path length is used also the measured absorbance should be 10 times smaller. For this reason the result should be 0,12.
Do you recommend anything to prevent contaminations by nucleic acids?Back to top
Nucleic acids could be removed easily with water. But it is also possible to clean the cuvette with lint-free paper tissue soaked with 6% sodium hypochlorite to remove nucleic acid contaminations.
In which direction I have to insert the Eppendorf µCuvette into the cuvette shaft?Back to top
There is no specific direction recommended. The Eppendorf µCuvette has to be placed in cuvette shaft, that the optical window is aligned with the direction of the light beam of Bio- or BioSpectrometer. But it is necessary that blank and sample are measured in the same direction.
Can the Eppendorf µCuvette be cleaned in an ultrasonic bath?Back to top
No, unfortunately we cannot recommend this.
Is it possible to use the Eppendorf µCuvette also in the BioPhotometer D30?Back to top
Yes, this is possible without any restrictions
What is the minimum volume for nucleic acid solution that I can use in the Eppendorf µCuvette that I can use in the BioPhotometer D30?Back to top
The smallest volume should be at least 1.5 µL.
What is the minimum volume for protein solution that I can use in the Eppendorf µCuvette that I can use in the BioPhotometer D30?Back to top
This depends also to some extend to on the proteins. We recommend a volume of at least 3 µL.
Is it possible to carry out a wavelength scan with the BioPhotometer D30?Back to top
No, this is unfortunately not possible. The BioPhotometer D30 is a fix wavelength photometer for specific methods.
What wavelength could be applied in the BioPhotometer D30?Back to top
The wavelengths 230, 260, 280, 320, 340, 405, 490, 562, 595 und 600 nm could be applied.
What methods are preprogrammed in the BioPhotometer D30?Back to top
Preprogrammed are all nucleic acid and protein detection methods and the OD 600 method for the determination of the optical density of bacterial or yeast cultures. Protein can be detected at 280 nm and via colorimetric assays like the BCA, the Bradford or the Lowry method.
Is it possible to use the BioPhotometer Data Transfer Software for the BioPhotometer D30?Back to top
No, this is unfortunately not possible. But this is also not necessary since the results are exported directly as Excel-files. An additional program for data management is therefore not necessary.
What can I do with the so called "impurity scan"?Back to top
With the "impurity scan" the absorbance of a certain wavelength range is shown for nucleic acid and protein measurements. With this function impurities in the sample can be easily detected via deviations from the known curve.
For what methods the "impurity scan" could be applied?Back to top
This function is available for nucleic acid and protein detections.
What wavelength range is used for the "impurity scan"?Back to top
For protein measurements it is the wavelength range from 245-350 nm and for nucleic acids the range from 225-350nm.
Is it possible to change the range for the "impurity scan" and can I display the single absorbance values of the "impurity scan"?Back to top
No, unfortunately both options are not possible. This can be done only in the BioSpectrometer
Can I use the Hellma® TrayCell with the BioPhotometer D30?Back to top
Yes, this is of course possible! Please check, that the cuvette with the required light beam hight of 8.5 mm is applied. Furthermore it is necessary for the precise calculation of the sample concentration that the correct light path length is set in the parameter section (stated on the lid)
What kind of cuvette path lengths could be applied for the BioPhotometer D30?Back to top
Like for the BioSpectrometer the cuvette path lengths of 10, 5, 2, 1, 0.2, 0.1 mm could be programmed.
Is there any possibility to check the BioPhotometer D30?Back to top
Yes, this is possible. We offer a calibration filter set especially for the BioPhotometer D30.
In what file-format are the files from the BioPhotometer D30 exported?Back to top
The data will be exported as Excel-files
Is it possible to select my data for the export?Back to top
Yes, this is possible! In the method you can select in the section "print&export" specific data packages and specific results for the export.
What kind of SI-units can be applied with the BioPhotometer D30?Back to top
This depends on the method. The free programmable methods offer 18 different kind of SI units. But it is possible at any time to edit new SI-Units?
In the BioPhotometer D30 certain factors are preprogrammed for the evaluation of specific proteins. Is it possible to apply factors also for other proteins?Back to top
Yes, this is possible. Select in the "parameter" section of the method "protein". With "edit" a calculation tools for protein factor starts. Here, you may set the factor for the new protein directly. It is also possible to calculate the factor via the absorbance coefficient and the molecular mass of the protein or via its value "A0.1%"
What is the meaning of the value "A0,1%"?Back to top
The meaning of A0,1% is the absorbance of the Protein with a concentration of 1mg/ml.
How many methods can be stored in the BioPhotometer D30?Back to top
About 100 methods can be stored in the BioPhotometer D30 or the BioSpectrometer.
How many results can be stored in the BioPhotometer D30?Back to top
About 1000 results can be stored in the BioPhotometer D30 or the BioSpectrometer.
 