"Clean rotor" appears on the display of my centrifuge. Do I now need to remove the rotor from the centrifuge, and, if yes, how?

This display appears every 200 centrifugation runs. For cleaning the rotor and the rotor chamber, the rotor must first be removed from the centrifuge. To do this, loosen the rotor nut by turning it counter-clockwise with the supplied rotor key. Please clean the rotor and rotor chamber only with mild cleaning agents (e.g. 70% alcohol solutions or commercially available alcoholic disinfectants).

"Does the Centrifuge 59xx have an energy saving function?“

"Err 8" appears on the display. What does this mean?

 
What is the maximum number of 15 mL conical tubes that can be centrifuged in the swing-bucket rotor A-4-38 of centrifuges 57xx?

A pop-up window asks me to perform a firmware update after I staretd my iPod. How can I perform the firmware update?

After adding the sample to the Eppendorf µCuvette no liquid column could be formed. What could be the reason for this?

Are 1.5 ml Safe-Lock micro test tubes safe for use in the food industry in Europe?

Are 1.5 ml Safe-Lock micro test tubes safe for use in the food industry in USA?

Are all mobile devices supported by VisioNize Lab Suite?

Are all parts that I need for the connection of the FemtoJet 4i/ 4x to the Eppendorf manipulators (e.g. InjectMan 4) included in the standard accessories or are additional parts required?