Can the CryoCube F101h be locked with a key?
Can the CryoCube F101h be locked with a key?
Yes, a lock is integrated in the door handle of the CryoCube F101h. Two keys are available.
Where do I find the control panel of the CryoCube F101h?
Where do I find the control panel of the CryoCube F101h?
The control panel of the CryoCube F101h is located in the upper right corner on front of the device.
What kind of insulation is used for the CryoCube F101h?
What kind of insulation is used for the CryoCube F101h?
The Freezer is insulated using polyurethane foam (PU) as well as vacuum insulation panels. This results in a low energy consumption of the F101h and leads to long warm-up time during power failure.
What means "Pull down time"?
What means "Pull down time"?
This is the time required for a unit to pull down from Room temperature to -80 °C. This usually takes between 3-4 hours for an Eppendorf freezer. The pull down data for Eppendorf freezers can be found in our familiy brochure "Pure. Ultracold".
How many boxes fit in an Eppendorf CryoCube F440n / F440h ?
How many boxes fit in an Eppendorf CryoCube F440n / F440h ?
​In one of the five compartments 4 racks will fit. The storage capacity is 20 racks per freezer, 16 boxes per rack and in total 320 x 2" boxes per freezer.
What is the design of the inner chamber of the CryoCube F440h and F440n ?
What is the design of the inner chamber of the CryoCube F440h and F440n ?
​The chamber of CryoCube F440h and F440n is composed of 5 compartments with 5 inner doors. It is made of stainless steel, which is easy to clean and disinfect.
What are the safety features of the CryoCube F440h/n ?
What are the safety features of the CryoCube F440h/n ?
​The handle can be locked with an integrated lock and additional padlock. A lockable metal plate protects the main and battery switches from accidental disconnection. All settings (alarms, delays and setpoints) can be protected by an individual code.
What is the pull down time, i.e. the time it takes for the CryoCube F440h/n (230 V/115V) to…
What is the pull down time, i.e. the time it takes for the CryoCube F440h/n (230 V/115V) to cool down from ambient to -80°C ?
​The pull-down time for the CryoCube F440h is 215 min for 230V and 180 min for 115V.
The pull-down time for the F440n is 3h 35 min for the 230 V variant and 3h for 115 V.
How does the insulation of the CryoCube F440h differ from that of the CryoCube F440n ? …
How does the insulation of the CryoCube F440h differ from that of the CryoCube F440n ?
​The insulation of the CryoCube F440h is composed of environmental friendly polyurethane foam and vacuum insulation panels. The CryoCube F440n is insulated exclusively with PU foam.
Are the racks of the U410 or U410HEF compatible with the CryoCube F440h/n ?
Are the racks of the U410 or U410HEF compatible with the CryoCube F440h/n ?
​Yes, the racks of the previous models fit into the CryoCube F440h/n. For more information, find the brochure "Store More" on our website.