Can you program multiple centrifugation steps in one program on the Centrifuge 5425? …
Can you program multiple centrifugation steps in one program on the Centrifuge 5425?
No, with one program you only set speed and time for one single centrifugation run.
Are the rotors exchangeable between the Centrifuge 5424 and 5425?
Are the rotors exchangeable between the Centrifuge 5424 and 5425?
No, unfortunately not.
Could you recommend what kind of divisible PCR plates should be used for the new PCR…
Could you recommend what kind of divisible PCR plates should be used for the new PCR swing-bucket rotor of the Centrifuge 5425?
Unskirted divisible PCR plates, like our Eppendorf twin.tec® PCR Plate 96, divisible must be used.
Can I use 0.1 mL PCR tubes for the new PCR swing-out rotor?
Can I use 0.1 mL PCR tubes for the new PCR swing-out rotor?
Yes, it is possible to use 0.1 mL and 0.2 mL PCR tubes or tube strips for this rotor.
Can the Eppendorf 5 mL Tubes with screw caps also be used for the 5 mL rotor of the…
Can the Eppendorf 5 mL Tubes with screw caps also be used for the 5 mL rotor of the Centrifuge 5425?
All Eppendorf 5 mL Tubes can be used in the 5 mL rotor of the Centrifuge 5425.
What kind of rotors will be offered for the Centrifuge 5425?
What kind of rotors will be offered for the Centrifuge 5425?
Rotor FA-24x2: Aerosol-tight with QuickLock® Lid for 24 x 1.5 / 2.0 mL tubes
Rotor FA-24x2 - PTFE: Aerosol-tight with QuickLock® Lid for 24 x 1.5 / 2.0 mL tubes, with PTFE coating for higher chemical resistance.
Rotor FA-10x5: Aerosol-tight with QuickLock® Lid for 10 x Eppendorf 5 mL tubes
Rotor FA-18x2 Kit: Aerosol-tight with QuickLock® Lid for 18 x 2 mL vessels incl. "Spin columns"
Rotor F-32x0.2-PCR: Rotor for 4 x PCR strips or 32 x PCR tubes
Rotor 96x0.2-PCR: Rotor for 12 x PCR strips (0.1 or 0.2 mL) or 4 x ¼ twin.tec PCR plate 96, unskirted divisible or 96 x 0.2 mL PCR Tubes.
How many protocols can be saved on the Centrifuge 5425?
How many protocols can be saved on the Centrifuge 5425?
Altogether 3 program slots are available on the Centrifuge 5425.
How can I calculate the RCF value from the RPM value for my protocols as precisely as…
How can I calculate the RCF value from the RPM value for my protocols as precisely as possible?
Use the menu function "ROTOR" for this purpose. By specifying the rotor and the vessel used, the required RCF value can be calculated exactly.
What is behind the "Finished Time" function of the Centrifuge 5425?
What is behind the "Finished Time" function of the Centrifuge 5425?
With the “Finished Time” function, you can see the time elapsed since the end of the centrifuge run.
What does the "Timer” function of the Centrifuge 5425 stand for?
What does the "Timer” function of the Centrifuge 5425 stand for?
With the "Timer" function, the start time can be delayed. This can be used for many nucleic acid purification kits where the buffer on the column should be incubated before the sample is centrifuged.
Are the rotors with QuickLock® lid of Centrifuge 5425 aerosol-tight?
Are the rotors with QuickLock® lid of Centrifuge 5425 aerosol-tight?
Yes, all rotors of the 5425 with QuickLock® lid are aerosol-tight.
Why are there two 24-place rotors for the Centrifuge 5425?
Why are there two 24-place rotors for the Centrifuge 5425?
The rotors have a different coating. The 24-place standard rotor FA-24x2 is with the anodized surface good resistant against chemicals and corrosion. The special rotor FA-24x2-PTFE with the PTFE coating is characterized by a even more higher resistance to chemicals.
Do you recommend a specific cleaning method for the rotors or the rotor chamber?
Do you recommend a specific cleaning method for the rotors or the rotor chamber?
Yes, please read chapter 6 of the Operating Manual or the White Paper No. 14 "Routine Maintenance of Centrifuges"
How should I lubricate the sealings of the QuickLock® rotors for the Centrifuge 5425? Is…
How should I lubricate the sealings of the QuickLock® rotors for the Centrifuge 5425? Is there a difference compared to former QuickLock® rotors?
Yes, we have improved our QuickLock® system for the Centrifuge 5425. Please lubricate the QuickLock® rotors of the Centrifuge 5425 as described in the manual or the rotor supplementary sheet.
Is it possible to switch off the automatic lid opening?
Is it possible to switch off the automatic lid opening?
Yes, you can switch off the automatic lid opening via the function “LID”.
What is the maximum rotational speed of the Centrifuge 5420?
What is the maximum rotational speed of the Centrifuge 5420?
The maximum speed for both fixed-angle rotors is 15,060 rpm. This corresponds to 21.300 x g for rotor FA-24x2 mL.
What is the smallest selectable speed with the Centrifuge 5420?
What is the smallest selectable speed with the Centrifuge 5420?
The smallest selectable speed is 100 rpm.
Is there a possibility to protect my programs on the Centrifuge 5420?
Is there a possibility to protect my programs on the Centrifuge 5420?
Yes, this is possible. You can activate the write-protection of the programs in the function menu.
How many programs can be saved on the Centrifuge 5420?
How many programs can be saved on the Centrifuge 5420?
Altogether 3 program slots are available on the Centrifuge 5420. They can be selected by pressing the program keys.
Is it possible to program multiple centrifugation steps in one program on the Centrifuge…
Is it possible to program multiple centrifugation steps in one program on the Centrifuge 5420?
No, with one program you only set speed and time for one single centrifugation run.
What kind of rotors will be offered for the Centrifuge 5420?
What kind of rotors will be offered for the Centrifuge 5420?
Rotor FA-24x2: Aerosol-tight for 24 x 1.5 / 2.0 mL tubes
Rotor F-32x0.2-PCR: Rotor for 4 x PCR strips or 32 x 0.2 mL PCR tubes
How can I calculate the RCF value from the RPM value for my protocols as precisely as…
How can I calculate the RCF value from the RPM value for my protocols as precisely as possible?
Use the menu function "ROTOR" for this purpose. By specifying the rotor and the vessel used, the required RCF value can be calculated exactly.
Is it possible to do a short run centrifugation on Centrifuge 5420?
Is it possible to do a short run centrifugation on Centrifuge 5420?
To start a short spin centrifugation, press or press and hold the short key.
Pressing and holding the short key: centrifuge runs for as long as the short key is pressed.
Briefly pressing the short key: centrifuge accelerates up to the set speed and then decelerates.
Is it possible to switch off the automatic lid opening on Centrifuge 5420?
Is it possible to switch off the automatic lid opening on Centrifuge 5420?
Yes, you can switch off the automatic lid opening via the function “LID”.
Do you recommend a specific cleaning method for the rotors or the rotor chamber of…
Do you recommend a specific cleaning method for the rotors or the rotor chamber of Centrifuge 5420?
Yes, please read chapter 6 of the Operating Manual or the White Paper No. 14 "Routine Maintenance of Centrifuges"
How should I lubricate the sealing of the FA-24x2 rotor for the Centrifuge 5420?
How should I lubricate the sealing of the FA-24x2 rotor for the Centrifuge 5420?
Yes, we have improved our QuickLock® system for the Centrifuge 5420. Please lubricate the QuickLock® rotors of the Centrifuge 5420 as described in the manual or the rotor supplementary sheet.
Why are the standard rotors in the micro centrifuges 5418 R, 5420, 5424 R, 5425, 5427 R,…
Why are the standard rotors in the micro centrifuges 5418 R, 5420, 5424 R, 5425, 5427 R, 5430 and 5430 R aerosol-tight?
At Eppendorf, the safety of the user always has the highest priority. That is why we now fit our centrifuges with aerosol-tight rotors as standard – at no extra charge.
What is the meaning of the red knob on the rotor lid?
What is the meaning of the red knob on the rotor lid?
A fixed-angle rotor with a red knob on the rotor lid indicates aerosol-tightness.
Are the rotors exchangeable between the Centrifuge 5424, 5425 and 5420?
Are the rotors exchangeable between the Centrifuge 5424, 5425 and 5420?
No, unfortunately not.
What is the UVette made of?
What is the UVette made of?
The UVette is made of a UV-transparent plastic.