How big is the g-force difference between the lowest and the highest plate when loading a…
How big is the g-force difference between the lowest and the highest plate when loading a bucket of the rotor A-2-DWP with four microplates?
The following g-forces are reached with the rotor A-2-DWP:
Lowest plate: 2,250 x g
Highest plate: 1,825 x g
Is there an adapter to spin microscope slides in centrifuge 5804 / 5804 R / 5810 / 5810 R? …
Is there an adapter to spin microscope slides in centrifuge 5804 / 5804 R / 5810 / 5810 R?
Up to 12 slides can be inserted into the CombiSlide adapter 5825706005 and then centrifuged in the swing out rotors of Centrifuge 5804/ 5804 R or 5810/ 5810 R.
Can 4 microplate buckets (order number 5810702003) be fitted into the rotor A-4-62…
Can 4 microplate buckets (order number 5810702003) be fitted into the rotor A-4-62 (5810709008) of Centrifuge 5810 / 5810 R?
Yes. It is the same rotor cross. Please note that mixed loading of the rotor with rectangular buckets and MTP buckets is not permitted.
Can rotor A-4-81 or S-4-104 be operated with rectangular buckets and MTP / Flex buckets? …
Can rotor A-4-81 or S-4-104 be operated with rectangular buckets and MTP / Flex buckets?
Yes, this so-called mixed loading is possible. Please ensure that the same type of buckets are always used in opposing positions.
Do the adapters for the 0.2/ 0.4/ 0.5/ 0.6 mL tubes, which are designed for the rotors…
Do the adapters for the 0.2/ 0.4/ 0.5/ 0.6 mL tubes, which are designed for the rotors F-45-30-11 and FA-45-30-11, also fit into the rotor F-34-6-38 with the adapter 5804 770.005 (for 1.5/ 2.0 mL tubes)?
Yes, these adapters can also be used in the rotor F-34-6-38 with the adapter 5804770005.
There are 2 different adapters for the 15 mL tubes. One for tubes with round bottoms and…
There are 2 different adapters for the 15 mL tubes. One for tubes with round bottoms and one with a conical bottom for conical tubes. Can I not use one adapter for both tubes?
No, unfortunately this is not possible. In order to be able to reach the high rotational speeds, the adapters must be form-locking.
Do the adapters for the 0.5 mL tubes fit into the rack for 1.5/ 2.0 mL tubes of the rotor…
Do the adapters for the 0.5 mL tubes fit into the rack for 1.5/ 2.0 mL tubes of the rotor T-60-11 for centrifuge 58xx?
Yes, the adapters can be inserted into the rotor for centrifugation of 0.5 mL tubes.
The writing on one rotor of the Centrifuge 5804/ 5804 R or 5810/ 5810 R says, for example,…
The writing on one rotor of the Centrifuge 5804/ 5804 R or 5810/ 5810 R says, for example, "6x125g". Does this mean that the centrifuge only enables an RCF of 125 or that it can only be loaded with a maximum of 6 times 125 gram?
This label states that the rotor can be loaded with a maximum of 6 x 125 gram. "g" is the official abbreviation for "gram". Where "g" is used as an abbreviation of "g-force", the number and "x g" are used. Example: 10,800 x g.
How many times can aluminum rotors be autoclaved?
How many times can aluminum rotors be autoclaved?
Aluminum rotors can be autoclaved up to 100 times for 20 minutes at 121 °C.
Does the Centrifuge 58xx R has an energy saving function?
Does the Centrifuge 58xx R has an energy saving function?
Yes, the ECO shut-off automatically switchs off the compressor after 8 hours of non-use. The device has a stand-by button.
Can the new rotor generation be used with older Centrifuges without retrofitting?
Can the new rotor generation be used with older Centrifuges without retrofitting?
Yes, you can use the new rotors from software version 8.4 onwards. From versionb 7.0 on a software update to version 8.4 can easily be performed by trained service technician. In case of older software versions, the complete PCB control/frequency converter must be replaced. You can see the current software version of your centrifuge briefly on the display after switching it on.
How does the Centrifuge recognize that a different rotor is in use?
How does the Centrifuge recognize that a different rotor is in use?
A new rotor automatically will be recognized as soon as it starts moving. Moving and recognition of the rotor can be performed manually by making it spin counterclockwise (roughly 200 rpm necessary), or by turning the Centrifuge on. As soon as the rotor starts spinning it will be recognized. When indicated, the Centrifuge shuts down under specification of max. speed in case the indicated max. speed of the previous rotor was higher. In this case the Centrifuge must be restarted.
Is it possible, in addition to 25 cm2 flasks, to use 75 cm2 flasks also?
Is it possible, in addition to 25 cm2 flasks, to use 75 cm2 flasks also?
Only a limited choice of 72 cm2 flasks are possible to be centrifuged: We recommend to use Saarstedt flasks (Article number 831811-002) with the cell culture flask adapter (Article number 5825719000) in combination with rotor A-4-81-MTP/Flex or A-4-62-MTP and Centrifuge 5810/ R.
Can the Centrifuge 5427 R be opened in the event of a power failure?
Can the Centrifuge 5427 R be opened in the event of a power failure?
Yes, in the event of a power failure, the unit can be opened by means of the rotor key. Before the centrifuge is opened, always remove the mains plug from the plug socket and wait for the rotor to come to a standstill. For the emergency release of the centrifuge lid, the small plastic cap on the right front corner of the 5427 R has to be removed. Then insert the rotor key into the nut underneath the cap and open the lid latches by turning in anticlockwise direction.
How can the Centrifuge 5427 R be cleaned?
How can the Centrifuge 5427 R be cleaned?
The Centrifuge 5427 R can be cleaned with neutral detergents. For disinfection, we recommend alcohol or alcoholic disinfectants. Make sure that the relevant legal requirements and directives are observed when choosing the disinfection method.
Sometimes my Centrifuge 5427 R displays "EP". What does that mean?"
Sometimes my Centrifuge 5427 R displays "EP". What does that mean?"
EP" is displayed when the automatic standby function is active. If the centrifuge remains unused for more than 15 min., the standby function is activated, indicated by "EP" on the display. By using any key, the centrifuge is re-activated and ready to use.
Can the Centrifuge 5427 R automatically recognize an installed rotor?
Can the Centrifuge 5427 R automatically recognize an installed rotor?
The Centrifuge 5427 R has an automatic rotor recognition system. A newly installed rotor is recognized at the beginning of the run. The maximum speed of each rotor is automatically limited, if necessary, to prevent over-speeding of the rotors.
What benefits do I gain from the centrifugation of my samples with the swing bucket rotor? …
What benefits do I gain from the centrifugation of my samples with the swing bucket rotor?
One advantage of centrifugation with the swing bucket rotor is the possibility of sedimenting molecules that cannot be pelletized with a fixed-angle rotor (Microgradient, e.g. blood / Ficoll).
What is the benefit of a so-called Kit rotor?
What is the benefit of a so-called Kit rotor?
When working with nucleic acid purification kits with spin columns by certain manufacturers, sometimes tube lids cannot be closed any more. This means you have to centrifuge with open tube lids. These tube lids might rip off during the stress of centrifugation and might damage the centrifuge. In order to prevent this, the kit rotor with the raised rotor rim was designed. Open tube lids, supported against the rim, will not rip off during centrifugation. Another feature of this rotor is the enlarged distance between the rotor bores. The enlarged distance between the bores offers a very convenient handling when loading / unloading the rotor. Especially when working with long spin columns (e. g. from Princeton Separations) you will need a larger distance between the bores to load the rotor fully and make use of the complete rotor capacity.
What tubes can be centrifuged with the Centrifuge 5427 R?
What tubes can be centrifuged with the Centrifuge 5427 R?
Due to the different rotors and adapters, a wide variety of tubes can be centrifuged with the
Centrifuge 5427 R:
• Micro test tubes (0.5 - 5 mL)
• PCR strips and 0.2 mL PCR tubes
• Spin columns (1.5/2.0 mL)
• Microtainer (0.6 mL
In which rotors can I centrifuge 0.2 mL PCR tubes?
In which rotors can I centrifuge 0.2 mL PCR tubes?
Up to 48 0.2 mL PCR tubes (or 6 x 8-tube PCR strips) can be centrifuged in the PCR rotor
F-45-48-5-PCR. Furthermore, 0.2 mL PCR tubes can be used in each rotor for 1.5/ 2.0 mL micro test tubes using an adapter
In which rotors can I centrifuge 0.5 mL micro test tubes?
In which rotors can I centrifuge 0.5 mL micro test tubes?
0.5 mL micro test tubes can be centrifuged in all rotors for 1.5/ 2.0 mL micro test tubes using an adapter.
In which rotors can I centrifuge 5 mL micro test tubes?
In which rotors can I centrifuge 5 mL micro test tubes?
Up to 16x 5 mL micro test tubes can be centrifuged in the aerosol-tight rotor FA-45-12-17.
In which rotor can I centrifuge spin columns?
In which rotor can I centrifuge spin columns?
Spin columns or small filtration columns, for example for centrifugation of DNA or RNA, can be used in all rotors for 1.5/2.0 mL micro test tubes.
In particular, however, the kit rotor F-45-24-11-Kit is specially suitable for such spin columns, as open tube lids are supported by the raised edge of this rotor type and are no longer torn off (e.g. during elution of the sample from the column into a micro test tube)
What is "Fast Temp"?
What is "Fast Temp"?
The "Fast Temp" function is a temperature control run of the Centrifuge 5427 R without samples and with a rotor and temperature-specific rotational speed in order to quickly bring the rotor chamber, incl. the rotor, buckets and adapters, to the set nominal temperature. In order that the samples can be centrifuged from the beginning on at the desired temperature, one starts this temperature control run during sample preparation by pressing the "Fast Temp" key. The display shows the current temperature and the g-force (rcf) / rotational speed (rpm) as well as "FT" as symbol for a "Fast Temp" temperature run. You can find a more detailed description in the operating instructions of the Centrifuge 5427 R.
When does the continuous cooling of the rotor chamber start?
When does the continuous cooling of the rotor chamber start?
The continuous cooling starts with each rotor standstill of the Centrifuge 5427 R in order to maintain the preselected nominal temperature of the rotor chamber not only during centrifugation. To this purpose the following prerequisites must be met:
- The centrifuge is switched on.
- The centrifuge lid is closed.
- The nominal temperature is lower than the ambient temperature.
- The centrifuge is not in standby mode.
Continuous cooling is ended by opening the centrifuge lid or by pressing any key.
What is the lowest temperature that can be reached with the Centrifuge 5427 R?
What is the lowest temperature that can be reached with the Centrifuge 5427 R?
The lowest temperature of the Centrifuge 5430 R is -11 °C. The device is thereby optimized with all rotors in such a way that at least 4 °C is maintained with all rotors at the appropriate maximum speeds.
What is the temperature range of the Centrifuge 5427 R?
What is the temperature range of the Centrifuge 5427 R?
With the Centrifuge 5430 R it is possible to set the temperature within a range from -11 °C to +40 °C.
Does the Centrifuge 5427 R has an energy saving function?
Does the Centrifuge 5427 R has an energy saving function?
Continuous cooling is switched off if the centrifuge is not used for longer than the preset time. The centrifuge switches to standby mode.
- Default settings: Continuous cooling ends after 8 h.
- Continuous cooling can be limited to 1h, 2h or 4 h
What is the SOFT function?
What is the SOFT function?
The SOFT function of the Centrifuge 5427 R allows slower and gentler acceleration and braking compared to the normal acceleration and braking times. As a result, delicate samples are not stressed unnecessarily. In addition, the risk of the pellets being shaken up again during braking is minimized.