FAQ

どちらの製品の情報をお探しですか?

製品群を選択してください。

 

その他の質問についてはアプリケーションサポート窓口までご相談ください。

設定されたフィルターはありません。
1206 ドキュメント
How many speed-controlled pumps do the DASGIP bioreactors feature?
How many speed-controlled pumps do the DASGIP bioreactors feature?
Up to eight peristaltic speed-controlled pumps per vessel offer continuous flow rates from 0.3 mL/h - 5 L/h. Each pump can be operated independently. Two different pump head sizes for the multi pump modules DASGIP MP4 and MP8, fitting various tube sizes make the DASGIP multi pumps a flexible and accurate solution.
How are DASGIP vessels agitated?
How are DASGIP vessels agitated?
DASGIP bioreactors feature powerful overhead drives with agitation speeds ranging 30 – 1,600 rpm. The DASGIP TC4SC4 controls temperature and agitation in up to four bioreactors. Users can individually define each outlet set-point as well as choose from different operation modes (e.g. continuous or interval agitation).
Do stainless steel components bear any risk of corrosion?
Do stainless steel components bear any risk of corrosion?
As it is highly resistant to most laboratory chemicals, stainless steel, e.g. 1.4571 (AISI-316Ti), 1.4404 (AISI-316L), is used for wetted parts in DASGIP Bioreactor Systems. A passive coating made of chrome oxide protects the steel from oxidizing agents. Despite these protective properties, corrosion can still occur in extreme situations such as mechanical surface damage or when the parts come in contact with concentrated acids or alkaline solutions. An indication of corrosion can be discoloration that is visible even after a thorough cleaning (ultrasonic bath with an approved stainless steel cleaning solution).
How long can the power interruption be before the cycler would not automatically restart? …
How long can the power interruption be before the cycler would not automatically restart?
The user can define the duration of interruption that the cycler would restart (up to 99min:59sec) under the “Cycler system function” (System node -- restart). The cycler then would not restart if the duration is longer than specified by user or if the interruption is less or equal 100min.
Does the Restart after a power interruption continue from where it left off, or does the…
Does the Restart after a power interruption continue from where it left off, or does the whole protocol restart from the beginning?
As long as the restart procedure is within the programmed restart time, the protocol will continue from the last step where it left off. For e.g. if the cycler was interrupted in the middle of step 20 (the 19th second of 60°C/20s), the cycler would restart from the beginning of step 20.
What is the max sample volume that I can use in a aluminum block?
What is the max sample volume that I can use in a aluminum block?
For PCR plates and 0.1/0.2mL tubes: 70µL. For 0.5mL PCR tubes: 100µL
What is the max sample volume that I can use in a silver block?
What is the max sample volume that I can use in a silver block?
For PCR plates and 0.1/0.2mL tubes: 100µL.
Can any of the Mastercycler nexus eco models be connected to any of the Mastercycler master…
Can any of the Mastercycler nexus eco models be connected to any of the Mastercycler master devices regardless of their specifications?
A Mastercycler nexus eco can only be connect to any Mastercycler nexus control unit. Mastercycler pro units can only be connected to other Mastercycler pro units. The two different models are not interchangeable.
What if I forgot my User password?
What if I forgot my User password?
Your Administrator can reset a new password for any User accounts using their Admin account.
What if I forgot my Administrator password?
What if I forgot my Administrator password?
Please contact your regional Eppendorf Service/Sales technician.
What is the distance between the closed flexlid and the block of Mastercycler nexus flat? …
What is the distance between the closed flexlid and the block of Mastercycler nexus flat?
7.5 mm
What is the block dimension of the Mastercycler nexus flat?
What is the block dimension of the Mastercycler nexus flat?
The wide of the black plate is 69 mm and the wide of the complete block is 82 mm. The length of the black part is 107 mm and the length of the complete block is 122 mm.
What is the max sample volume that I can use in the 96-well block devices?
What is the max sample volume that I can use in the 96-well block devices?
100 µL (for both aluminum and silver blocks)
What is the max sample volume that I can use in the 384-well block device?
What is the max sample volume that I can use in the 384-well block device?
25 µL
Can I transfer data (e.g. user profiles, protocols) from my older Eppendorf thermal cyclers…
Can I transfer data (e.g. user profiles, protocols) from my older Eppendorf thermal cyclers (Mastercycler gradient, nexus or pro families) to the new Mastercycler X50?
As the Mastercycler uses modern user interface and software structure, transfer of data from older Eppendorf Mastercyclers is currently not possible.
Can the Mastercycler X50 be connected to Mastercycler pro devices in a network?
Can the Mastercycler X50 be connected to Mastercycler pro devices in a network?
Networking is only possible within members of the Mastercycler X50 family. Up to 10 Mastercycler X50 devices can be connected in a device-controlled network while up to 50 Mastercycler X50 can be connected in a PC-controlled network.
What is 2D-Gradient?
What is 2D-Gradient?
2D-Gradient is a 2-dimensional gradient technology that allows two gradients in the same run. This allows you to optimize the annealing and the denaturation temperatures in parallel in a single run, thus saving valuable optimization time.
How can I connect a master device and an eco device of the Mastercycler X50?
How can I connect a master device and an eco device of the Mastercycler X50?
One eco device can be connected to a master device directly via an Ethernet cable between the "Cycler ports". If more than one eco devices are to be connected, an Ethernet switch is required. Each of the master and eco devices should be connected to the Ethernet switch via Ethernet cables from the "Cycler ports".
Does the Mastercycler X50 has password protection function for users?
Does the Mastercycler X50 has password protection function for users?
Yes, the Mastercycler X50 has extensive User Management functions with three level of user roles: Administrator, User with standard rights and User with restricted rights. The user accounts can be protected via Password or PIN system.
How fast is the silver block of Mastercycler X50?
How fast is the silver block of Mastercycler X50?
The block has heating rate of 10°C/s and cooling rate of 5°C/s.
What is the temperature accuracy of the block?
What is the temperature accuracy of the block?
±0.15°C
What is the temperature homogeneity of the block?
What is the temperature homogeneity of the block?
at 20-72°C: ±0.2°C
at 72-95°C: ±0.3°C
Does the Mastercycler X50 has in-device self-test function?
Does the Mastercycler X50 has in-device self-test function?
Yes. No other accessories are required.
Where can I find license information about software components of the Mastercycler X50? …
Where can I find license information about software components of the Mastercycler X50?
The software included in this product contains copyrighted software that is licensed under the GPL. A copy of that license is included in the settings of the device under “About this Mastercycler X50”. You may obtain the complete corresponding source code from us for a period of three years after our last shipment of this product. Please direct your request to eppendorf@eppendorf.com. This offer is valid to anyone in receipt of this information.
What is required to set up a complete network?
What is required to set up a complete network?
The minimum installation comprises one PC, 1 unit (up to 50 units) X50 without touchscreen interface, and Ethernet cable. When connecting more than 1 Mastercycler X50, an Ethernet hub and an additional Ethernet cable (for the PC to ethernet hub connection) is required.
What is the PC requirement?
What is the PC requirement?
CycleManager server
- Windows 7 Pro/Enterprise, Windows 10 Pro/Enterprise
- 4 GB RAM
- 2 TB of free hard disk space
- Intel Core i5, 2.8 GHz clock frequency (or higher)
- 2 Ethernet ports

Cycler Bridge
- Windows 7 Pro/Enterprise, Windows 10 Pro/Enterprise
- 1 GB RAM
- 500 GB of free hard disk space
- Intel Core i5, 2.8 GHz clock frequency (or higher)
- 2 Ethernet ports

Client
- Internet browser Chrome
- Screen resolution 1366 × 768 pixels (or higher)
Which Ethernet Hubs can be used for the software?
Which Ethernet Hubs can be used for the software?
We recommend 100 Mbit to 1 Gbit connections with auto-negotiation.
What are the IT infrastructure requirements?
What are the IT infrastructure requirements?
All X50 devices must be in the same network as the server installation.
There are 4 possible installation options:
1. Independent PC-cycler network
2. Local server network
3. Cyclers located at different locations (subnets)
4. Cyclers located at different locations + IT server
For more info, please contact your local sales representative.
Which ports need to be whitelisted for the cyclemanager to work?
Which ports need to be whitelisted for the cyclemanager to work?
1. Port 3030 needs to be accessible on the CycleManager X50 server from the client PC(s).
2. Port XXXX needs to be accessible on the bridge PC(s) from the CycleManager X50 server.
How many CycleManager X50 licenses are provided?
How many CycleManager X50 licenses are provided?
One. Since the server is accessed via Google Chrome client, multiple licenses are not needed for multiple users to access the software.