- Classic interface
- Volume: 740 L (26.1 ft³)
- Consumes only 11.8 kWh/day (green; -80 °C; 230 V))
- Holds up to 57,600 samples
- Pull-down time to -80 °C: 4 h 10 min
- 5 years warranty incl. parts and labor
- Polyurethane foam + Vacuum insulation panels (body)
- 3 compartments with 3 insulated and sealed inner doors
- Left-handed outer door, air-cooled
- Compatible with CO2/LN2 injection system (Type "narrow") as backup for highest sample safety
CryoCube® F740h
Product Information
- 740 L, classic interface, VIP, hydrocarbon cooling, air cooling
- handle left side, 3 compartments, 120 – 208 V/60 Hz (US)
-
Protect and Track your samples with our SafeCode barcoded cryovials and tubes
-
Find the right racks for your freezer (change browser if necessary)
You will find additional download material at the bottom of this page
Technical Data
| Specifications | CryoCube® F740h, 3c. |
|---|
|
Box capacity per freezer |
|
| 53 mm (2 in) tall boxes | 576 |
| 64 mm (2.5 in) tall boxes | 432 |
| 76 mm (3 in) tall boxes | 360 |
| 102 mm (4 in) tall boxes | 288 |
| 127 mm (5 in) tall boxes | 216 |
|
DWP capacity per freezer |
|
| 102 mm (4 in) tall boxes | 864 |
|
Box capacity per rack |
|
| 53 mm (2 in) tall boxes | 32 |
| 64 mm (2.5 in) tall boxes | 24 |
| 76 mm (3 in) tall boxes | 20 |
| 102 mm (4 in) tall boxes | 16 |
| 127 mm (5 in) tall boxes | 12 |
| DWP | 48 |
| Cooling liquid | hydrocarbon cooling (R290 / R170) |
| Heat output | 492 W |
| Capacity | 740 L (26.1 ft³) |
| No. of compartments | 3 |
| No. of internal doors | 3 |
| Max. racks per freezer | 18 |
| Max. power consumption | 11800 W |
| Average Daily Consumption (230 V) | 11.8 kWh/day |
| Average Daily Consumption (208 V) | 11.8 kWh/day |
| Average Daily Consumption (115 V) | 13.0 kWh/day |
| Dimensions (W × D × H) | 111 × 91.3 × 197.3 cm / 43.7 × 36 × 77.7 in1)2)3) |
| Shipping weight | 376 kg / 829 lb |
| Product weight | 326 kg / 719 lb |
| Access port | 2 |
| Pull-down time to -80 °C | 4 h 10 min4) |
1)Optional CO2/LN2 backup systems add 8.65 cm/3.5 in to height.
2)To allow for handles and hinges, add 80 mm to width of upright freezers and 110 mm to the depth of chest freezers.
3)Door open adds up to 15 cm.
4)Empty freezer with shelves fitted, upright freezers only, pull down from 21 – 23 °C ambient conditions.