text.skipToContent text.skipToNavigation
Поиск результатов не дал
Варианты поиска
Семейства изделий
Show all Product Families ()
search.autocomplete.variants.headline
Show all Articles ()
Содержание
Show all Content ()

What are the first steps after a freezer alarm?

Lab Academy

Ask the expert - Answer by Dr. Jan Bebermeier, Global Marketing Manager Cold Storage at Eppendorf

This article appeared first in “ Inside Cell Culture ”, the monthly newsletter for cell culture

A freezer alarm requires the existence of a freezer alarm system. In principle, there are two ways to be informed about a freezer/ power failure. The classic way is a call by the security desk about some beeping or flashing freezer in the laboratory or a hallway. Normally, you do not get any further information about the situation. The second way of information about an alarm is a personal message on your smartphone about a freezer alarm. This type of alarm is based on a cloud-based monitoring and notification system. Starting from the alarm, keep calm: You have several hours after the first alarm until the temperature inside the freezer is at a critical level. Based on data from an external test house, Eppendorf CryoCube freezers like the F440h or F570h with a 2/3 filling require ca. 10 hours to (passively) warm-up their chamber temperature after power failure from -80°C to -50°C. If the alarm is notified via a convenient up-to-date digital system, you can log-in from home or wherever you are when the alarm happened. The real situation can be pre-checked like temperature within the freezer, time of failure, indications about the causes,… When you finally decide to go to the laboratory, there are different aspects to consider:

  • Keep the freezer door closed until you really start the sample evacuation - everyone checking the samples does not help, it only speeds up the warming inside the freezer.
  • Check the power supply (power cable at instrument side as well as at socket side), try to re-boot the freezer, change also the socket/ power line and try to restart the instrument. If this also fails, continue:
    • Use your “Sample evacuation plan”: Where are free locations in other freezers in the building for complete racks or boxes with high value samples? If you are lucky, there is a back-up freezer in the building
    • Organize ice when transportation of super-critical samples takes longer than a few minutes
    • After taking care of the samples, call the service hotline of the freezer supplier