- Mechanical air-cushion pipette with standard 3-button operation (one operating element for volume selection, one for pipetting, and one for tip ejection)
- Ideal for accurate, easy, and ergonomic manual pipetting of aqueous solutions
- Feel the difference in weight and pipetting forces: Designed according to the Eppendorf PhysioCare Concept® principles
- Benefit from one of the lowest operating and tip ejection forces on the pipette market, helping to reduce hand and arm strain
- Enjoy minimal tip attachment forces due to a spring-loaded tip cone (available for all pipettes up to 1 mL)
- Temporarily adjust your pipette in seconds for better accuracy when pipetting challenging liquids like ethanol or when pipetting at high altitudes. Return to factory adjustment without calibration
- Autoclave the entire pipette or only the lower part according to your needs and applications to ensure a proper decontamination
- Premium material ensures a high UV and chemical resistance
- With the channel display, you can always use the pipette in the same position
- Increase your flexibility and remove individual channels as necessary
Eppendorf Research® plus
Product Information
- 12-channel
- 0,5 – 10 µL, medium gray, variable volume
- incl. epT.I.P.S.® Box 2.0 with 96 pipette tips
You will find additional download material at the bottom of this page
Pipette Tips (30)
Services (8)
Accessories (6)
Spare Parts (3)
Price
820,00 SEK
|
Price
800,00 SEK
|
Price
910,00 SEK
|
Price
2 330,00 SEK
|
Price
840,00 SEK
|
Price
1 090,00 SEK
|
Price
400,00 SEK
|
Price
500,00 SEK
|
Price
750,00 SEK
|
Price
1 000,00 SEK
|
Technical Data
Features
Technical Data
Specifications | Eppendorf Research® plus - 3125000028 |
---|
ACT label | − |
Product type | Pipettes |
Pipetting type | Air-cushion system |
Number of channels | 12-channel |
Volume range | 0,5 – 10 µL |
Color code | medium gray |
Volume increment | 0.01 µL |
Operating mode | mechanical |
Spring-loaded tip cone | yes |
Random measurement error1) |
0.5 µL: ± 8%; ± 0.04 µL 1 µL: ± 5%; ± 0.05 µL 5 µL: ± 2%; ± 0.1 µL 10 µL: ± 1%; ± 0.1 µL |
Systematic measurement error1) |
0.5 µL: ± 12%; ± 0.06 µL 1 µL: ± 8%; ± 0.08 µL 5 µL: ± 4%; ± 0.2 µL 10 µL: ± 2%; ± 0.2 µL |
Autoclavable | yes |
Consumables | epT.I.P.S.® |
Open system (third party consumables) | yes |
Delivery content | Pipette, epT.I.P.S.® Box 2.0 pipette tips, Operating manual, Certificate, Adjustment tool (Allen key with blue handle), Adjustment seal for temporary adjustment (x5), Grease for pipettes, Locking clip (to prevent spring action) (x3), O-ring tool |
1)The listed values for 10%, 50%, and 100% of the nominal volume are based on tests complying with DIN EN ISO 8655, Parts 2 or 5. If listed, the smallest adjustable volume provides additional data.