- Volume class: 570 L (20.1 ft3), e.g. up to 40,000 samples
- Efficient insulation, composed of traditional insulation ("n"-version) and vacuum insulation panels ("h"-version), saves energy
- Efficient compressor control system reduces cycle times to lower energy consumption and increases freezer longevity
- Automatic vent port on front door of upright CryoCube models improves energy consumption and enables fast sample access
- Lockable main switch to prevent switch-offs by accident
- Location for chart recorder for data documentation
- Ergonomic door handle of CryoCube models based on Eppendorf PhysioCare Concept® requires less force and provides easier access
- Air-cooled as well as water-cooled versions for further energy saving
- Stainless steel interior for easy cleaning during maintenance and defrost cycles; powder-coated steel exterior
CryoCube® F570h
Product Information
- 570 L, classic interface, VIP, green cooling liquids, air cooling
- handle left side, 5 compartments, ACT, 230 V/50 Hz (EU/GB)
You will find additional download material at the bottom of this page
Technical Data
Features
Technical Data
Specifications | CryoCube® F570h |
---|
ACT label | yes |
Cooling type | air cooling |
Cooling liquid | green cooling liquids (R290 / R170) |
Heat output | 308 W |
Insulation | vacuum insulation paneling / polyurethane foam |
Capacity | 570 L (20,1 ft³) |
No. of compartments | 5 |
No. of internal doors | 5 |
Max. racks per freezer | 25 |
Noise level | 57.7 dBA |
High-Efficiency | yes |
Power supply | 230 V, 50 Hz |
Max. power consumption | 7400 W |
Average Daily Consumption (230 V) | 7.4 kWh/day |
Average Daily Consumption (208 V) | − |
Average Daily Consumption (115 V) | 7.1 kWh/day |
Dimensions (W × D × H) | 110,3 × 89,8 × 195,0 cm / 43,4 × 35,4 × 76,8 in1)2)3) |
Footprint (WxD) | 110.3 × 89.9 cm |
Shipping weight | 388 kg / 855 lb |
Product weight | 341 kg / 752 lb |
Temperature range | -50 °C to -86 °C |
Access port | 2 |
Remote Device Monitoring and Notifications (VisioNize®) | VisioNize® box enabled |
Pull-down time to -80 °C | 3 h 30 min4) |
Box capacity per freezer |
|
53 mm (2 in) tall boxes | 400 |
64 mm (2.5 in) tall boxes | 300 |
76 mm (3 in) tall boxes | 200 |
102 mm (4 in) tall boxes | 200 |
127 mm (5 in) tall boxes | 200 |
DWP capacity per freezer |
|
102 mm (4 in) tall boxes | 600 |
Box capacity per rack |
|
53 mm (2 in) tall boxes | 16 |
64 mm (2.5 in) tall boxes | 12 |
76 mm (3 in) tall boxes | 8 |
102 mm (4 in) tall boxes | 8 |
127 mm (5 in) tall boxes | 8 |
DWP | 24 |
Included |
|
CO2 backup system | no |
LN2 backup system | no |
Chart recorder | no |
1)Optional CO2/LN2 backup systems add 8.65 cm/3.5 in to height.
2)To allow for handles and hinges, add 80 mm to width of upright freezers and 110 mm to the depth of chest freezers.
3)Door open adds up to 15 cm.
4)Empty freezer with shelves fitted, upright freezers only, pull down from 21 – 23 °C ambient conditions.