- Mechanical air-cushion pipette with outstanding precision and accuracy - operated with just one button for volume selection, pipetting and tip ejection
- With its innovative one-button operation, the Reference 2 allows fast and ergonomic handling with reduced operating efforts and active aerosol reduction. A clear haptic feedback helps to avoid accidental tip ejection
- 4-digit display with magnifying window for quick and secure volume setting and identification (clearly visible from every angle), incl. volume lock
- Spring-loaded tip cones improve user-to-user reproducibility and ergonomics by lowering the tip attachment force
- The outstanding precision and accuracy of Reference 2 provides reliable results
- A temporary adjustment allows for an easy adjustment of your Reference 2 for the most accurate pipetting of different liquids or in changing external conditions without the need for a full calibration
- The Reference 2 multi-channel variants offer high flexibility incl. a spring-loaded tip cone which can be switched on/off optionally
Eppendorf Reference® 2
Product Information
- 8-channel
- 10 – 100 µL, yellow, variable volume
- incl. epT.I.P.S.® Box 2.0 with 96 pipette tips
-
Learn more about sustainability & ergonomics at Eppendorf
-
Download your manufacturer's certificate and operating manual here
Vous trouverez plus d’informations à télécharger au bas de cette page
Pipette Tips (32)
Services (11)
Accessories (30)
Spare Parts (7)
Réf. catalogue
0030078543
|
Réf. catalogue
0030078713
|
Réf. catalogue
0030000889
|
Réf. catalogue
0030076150
|
Réf. catalogue
0030000870
|
Réf. catalogue
0030078551
|
Réf. catalogue
0030078837
|
Réf. catalogue
0030078721
|
Réf. catalogue
0030073835
|
Réf. catalogue
0030075242
|
Technical Data
| Caractéristiques | Eppendorf Reference® 2 - 4926000034 |
|---|
| Type de produit | Pipettes |
| Type de pipetage | Air-cushion system |
| Nombre de canaux | 8-channel |
| Plage de volume | 10 – 100 µL |
| Color code | yellow |
| Incrément de volume | 0.1 µL |
| Mode de fonctionnement | mechanical |
| Embout porte-cône à amortisseur de force | yes |
| Erreur de mesure aléatoire1) |
10 µL: ± 2%; ± 0.2 µL 50 µL: ± 0.8%; ± 0.4 µL 100 µL: ± 0.3%; ± 0.3 µL |
| Erreur de mesure systématique1) |
10 µL: ± 3%; ± 0.3 µL 50 µL: ± 1%; ± 0.5 µL 100 µL: ± 0.8%; ± 0.8 µL |
| Autoclavable | yes |
| Consumables | epT.I.P.S.® |
| Système ouvert (consommables de tiers) | yes |
| Contenu de la livraison | Pipette, Operating manual, Certificate, Adjustment seal for temporary adjustment, Adjustment seal for factory adjustment (x2), Grease for pipettes, O-ring tool |
1)The listed values for 10%, 50%, and 100% of the nominal volume are based on tests complying with DIN EN ISO 8655, Parts 2 or 5. If listed, the smallest adjustable volume provides additional data.