- Mechanical air-cushion pipette with outstanding precision and accuracy - operated with just one button for volume selection, pipetting and tip ejection
- With its innovative one-button operation, the Reference 2 allows fast and ergonomic handling with reduced operating efforts and active aerosol reduction. A clear haptic feedback helps to avoid accidental tip ejection
- 4-digit display with magnifying window for quick and secure volume setting and identification (clearly visible from every angle), incl. volume lock
- Spring-loaded tip cones improve user-to-user reproducibility and ergonomics by lowering the tip attachment force
- The outstanding precision and accuracy of Reference 2 provides reliable results
- A temporary adjustment allows for an easy adjustment of your Reference 2 for the most accurate pipetting of different liquids or in changing external conditions without the need for a full calibration
- The Reference 2 multi-channel variants offer high flexibility incl. a spring-loaded tip cone which can be switched on/off optionally
Eppendorf Reference® 2
Product Information
- 8-channel
- 30 – 300 µL, orange, variable volume
- incl. epT.I.P.S.® Box 2.0 with 96 pipette tips
-
Learn more about sustainability & ergonomics at Eppendorf
-
Download your manufacturer's certificate and operating manual here
You will find additional download material at the bottom of this page
Pipette Tips (13)
epServices (13)
Accessories (30)
Spare Parts (7)
Catalog no.
0030000900
|
Catalog no.
0030073452
|
Catalog no.
0030073835
|
Catalog no.
0030076168
|
Catalog no.
0030076338
|
Catalog no.
0030075242
|
Catalog no.
0030000897
|
Catalog no.
0030078560
|
Catalog no.
0030078730
|
Catalog no.
0030075447
|
Technical Data
| Specifications | Eppendorf Reference® 2 |
|---|
| Tipo do produto | Pipettes |
| Tipo de pipetagem | Air-cushion system |
| Número de canais | 8-channel |
| Intervalo de volumes | 30 – 300 µL |
| Color code | orange |
| Incremento de volume | 0.2 µL |
| Modod e operação | mechanical |
| Cone de ponteira com mola | yes |
| Erro de medição aleatória1) |
30 µL: ± 1%; ± 0.3 µL 150 µL: ± 0.5%; ± 0.75 µL 300 µL: ± 0.3%; ± 0.9 µL |
| Erro de medição sistemática1) |
30 µL: ± 3%; ± 0.9 µL 150 µL: ± 1%; ± 1.5 µL 300 µL: ± 0.6%; ± 1.8 µL |
| Autoclavável | yes |
| Consumables | epT.I.P.S.® |
| Sistema aberto (consumíveis de terceiros) | yes |
| Conteúdo fornecido | Pipette, epT.I.P.S.® Box 2.0 pipette tips, Adjustment seal for temporary adjustment, Adjustment seal for factory adjustment (x2), Grease for pipettes, O-ring tool |
| Rótulo ACT | − |
1)The listed values for 10%, 50%, and 100% of the nominal volume are based on tests complying with DIN EN ISO 8655, Parts 2 or 5. If listed, the smallest adjustable volume provides additional data.