text.skipToContent text.skipToNavigation
No results found
Search Suggestions

Accessories (5)


    • empty box for 8 Combitips® advanced (0.1 – 10 mL)
    • Eppendorf Quality, 1 piece
    • Replaces order no. 0030089758
    More Product Details ...
    Catalog no. 0030089952

      • contains one Combitips® advanced sample of each size and Adapter advanced 25 and 50 mL
      • Eppendorf Quality
      More Product Details ...
      Catalog no. 0030089936

        • 1-channel
        • 1 µL – 10 mL
        • incl. 1x Combitips® advanced 2.5 mL (green) tip and Pipette Holder 2
        More Product Details ...
        Catalog no. 4982000012

          • 1-channel
          • 1 µL – 50 mL
          • incl. charging cable and Combitips® advanced assortment pack (1 tip of each size)
          More Product Details ...
          Catalog no. 4987000029

            • 1-channel
            • 1 µL – 50 mL
            • incl. charging cable and Combitips® advanced assortment pack (1 tip of each size)
            More Product Details ...
            Catalog no. 4987000010

            Technical Data

            Features

            Specifications Combitips® advanced - 0030089448
            ACT label yes 
            Product type Consumables 
            Pipetting type Positive displacement system 
            Tip size 2.5 mL 
            Color code green 
            Tip color colorless tips 
            Purity grade Eppendorf Quality 
            Packaging 100 pcs. (4 bags × 25 pcs.) 
            Open system (third party consumables) no 

            Performance data with Multipette E3 & E3x / Repeater E3 & E3x

            Random measurement error 25 µL: ± 3.5%; ± 0.875 µL
            250 µL: ± 0.45%; ± 1.125 µL
            1250 µL: ± 0.3%; ± 3.75 µL
            2500 µL: ± 0.15%; ± 3.75 µL 
            Systematic measurement error 25 µL: ± 2%; ± 0.5 µL
            250 µL: ± 0.8%; ± 2 µL
            1250 µL: ± 0.5%; ± 6.25 µL
            2500 µL: ± 0.5%; ± 12.5 µL 
            Volume range 25 – 2.500 µL  

            Performance data with Multipette M4 / Repeater M4

            Random measurement error 25 µL: ± 8%; ± 2 µL
            50 µL: ± 0.8%; ± 0.4 µL
            250 µL: ± 0.6%; ± 1.5 µL
            500 µL: ± 0.3%; ± 1.5 µL 
            Systematic measurement error 25 µL: ± 4%; ± 1 µL
            50 µL: ± 0.8%; ± 0.4 µL
            250 µL: ± 0.6%; ± 1.5 µL
            500 µL: ± 0.5%; ± 2.5 µL 
            Volume range 25 – 500 µL