Dispensing head 96-channel 5 – 1,000 µL

카탈로그 번호 5069000535

  • 96-channel, for epMotion® 96 Flex in hard shell case, needs base device for operation
  • 5 – 1,000 µL

이 페이지 하단에서 추가 다운로드 자료를찾을 수 있습니다.

(6)


    • with filter
    • PCR clean, 10 – 1,000 µL, ACT, 2,304 tips (24 trays × 96 tips)
    추가 제품 정보...
    카탈로그 번호 0030014510

      • without filter
      • Eppendorf Quality, 5 – 300 µL, ACT, 2,304 tips (24 trays × 96 tips)
      추가 제품 정보...
      카탈로그 번호 0030014464

        • with filter
        • PCR clean, 5 – 300 µL, ACT, 2,304 tips (24 trays × 96 tips)
        추가 제품 정보...
        카탈로그 번호 0030014472

          • without filter
          • Eppendorf Quality, 10 – 1,000 µL, ACT, 2,304 tips (24 trays × 96 tips)
          추가 제품 정보...
          카탈로그 번호 0030014502

            • with filter
            • PCR clean and sterile, 5 – 300 µL, ACT, 2,304 tips (24 trays × 96 tips)
            추가 제품 정보...
            카탈로그 번호 0030014537

              • with filter
              • PCR clean and sterile, 10 – 1,000 µL, ACT, 2,304 tips (24 trays × 96 tips)
              추가 제품 정보...
              카탈로그 번호 0030014570

              사양 Dispensing head 96-channel 5 – 1,000 µL
              Pipetting type Air-cushion system 
              Number of channels 96-channel 
              Application Cell culture, Molecular biology 
              Volume range 5 – 1,000 µL 
              Operating mode
              Autoclavable no 
              Power supply
              Consumables epT.I.P.S.® Motion Reload System 
              Max. power consumption
              Dimensions (W × D × H)
              Product weight 2.7 kg / 6.0 lb
               
              Testing volume 300 µL Tips
              5 µL
              30 µL
              150 µL
              300 µL
              1000 µL Tips
              10 µL
              100 µL
              500 µL
              1000 µL 
              Systematic measurement error (determined in pipette mode) 300 µL Tips
              ≤ 3 % (not guaranteed, individual adjustment of device and dispensing head may be required)
              ≤ 2.5 %
              ≤ 1.5 %
              ≤ 1.5 %
              1000 µL Tips
              ≤ 3 % (not guaranteed, individual adjustment of device and dispensing head may be required)
              ≤ 1.5 %
              ≤ 1.5 %
              ≤ 1 % 
              Random measurement error (determined in pipette mode) 300 µL Tips
              ≤ 2.5 % at 30 µL
              1000 µL Tips
              ≤ 1.5 % at 100 µL 