- Effective contamination protection: 140 °C sterilization routine
- Uniform atmosphere: Dual-channel IR-sensor (no TC-sensor), temperature uniformity verified (ISO 12880:2007-05), temperature and CO2 recovery after door opening < 5 min
- Easy cleaning: Stainless-steel chamber without seams, fan, ducts, covers etc. and removable water tray for safe and easy weekly cleaning and refill according to good cell culture practice
- Save money: Up to 25 % more usable space by fanless design, 3/4 shelves (max. load per shelf: 10 kg/22 lbs) + 2 access ports included, no costly internal expendable parts like fan-associated HEPA-filters that require bi-annual replacement
- Enhanced vibration and turbulence protection: No fan inside
- Low gas consumption: <1 L CO2/24 h (5 % CO2, w/o door opening)
- You need more? If you need more options (e.g. O2 control, humidity monitoring, copper interior), 180 °C disinfection, or future flexibility – consider the CellXpert® C170i
ZA | ZAR
ZA | ZAR
-
- Benchtop Centrifuges
- Floor-Standing Centrifuges
- Refrigerated Centrifuges
- Microcentrifuges
- Multipurpose Centrifuges
- High-Speed Centrifuges
- Ultracentrifuges
- Concentrator
- High-Speed and Ultracentrifugation Consumables
- Accessories
- Tubes
- Plates
- Device Management Software
- Sample and Information Management
You are about to leave this site.
Please be aware that your current cart is not saved yet and cannot be restored on the new site nor when you come back. If you want to save your cart please login in into your account.
myEppendorf
CellXpert® C170
Product Information
- non-segmented inner door, handle right side
- 220 – 240 V/50 – 60 Hz (EU), 3 shelves included
You will find additional download material at the bottom of this page
Technical Data
Features
Technical Data
Specifications | CellXpert® C170 |
---|
Capacity | 170 L (6.0 ft³) |
Power supply | 220 – 240 V, 50 – 60 Hz |
Dimensions (W × D × H) | 71.8 × 71.5 × 90.0 cm / 28.3 × 28.1 × 35.4 in |
Product weight | 113 kg / 249 lb |
High temperature disinfection (HTD) | 140 °C |
Sealed inner glass doors | yes |
Segmented door options | 4 segments |
Temperature control increment | ±0.1 °C |
Temperature stability at 37 °C | ±0.1 °C |
Temperature range | Ambient +4 °C to 50 °C |
Temperature uniformity | ±0.3 °C |
CO2 range | 0.1 – 20 % |
CO2 control increment | ±0.1 % |
CO2 stability at 5 % CO2 | ±0.1 % |
CO2 uniformity | ±0.1 % |
Number of shelves (standard/max) | 3 (4 with 4-segmented door)/8 |
Max. load/shelf | 10 kg |
Max. load/rack | 40 kg |
Data export interface | Ethernet |
Benchtop (B), Under Bench (U), Floor (F) or Stackable (S) | B, U, F, S (x2) |
Copper package | – |
Access port | 2 |
On-board data logging | – |
Water/Humidity monitoring | – |
BMS relay | yes |
Temperature range | Ambient +4 °C to 50 °C |
Temperature accuracy | ±0.4 °C |