MENU
US | USD
-
-
-
-
- Forum Labo 2025
- Advanced Therapies Week (ATW) 2025
- SLAS Europe 2025
- Bioprocessing Summit Europe 2025
- Medlab Middle East 2025
- SLAS International 2025
- Biologics World Nordics 2025
- ASIA LABEX: The Lab Show 2025
- BioProcess International Europe 2025
- ISEV 2025
- Future Labs Live 2025
- Cell 2025
- ASIA LABEX: The Lab Show 2025
- Stem Cell Community Day 2025
-
-
-
-
- Forum Labo 2025
- Advanced Therapies Week (ATW) 2025
- SLAS Europe 2025
- Bioprocessing Summit Europe 2025
- Medlab Middle East 2025
- SLAS International 2025
- Biologics World Nordics 2025
- ASIA LABEX: The Lab Show 2025
- BioProcess International Europe 2025
- ISEV 2025
- Future Labs Live 2025
- Cell 2025
- ASIA LABEX: The Lab Show 2025
- Stem Cell Community Day 2025
US | USD
-
- Benchtop Centrifuges
- Floor-Standing Centrifuges
- Refrigerated Centrifuges
- Microcentrifuges
- Multipurpose Centrifuges
- High-Speed Centrifuges
- Ultracentrifuges
- Concentrator
- High-Speed and Ultracentrifuge Consumables
- Centrifuge Tubes
- Centrifuge Plates
- Device Management Software
- Sample and Information Management
-
- All Pipettes, Dispensers & Automated Liquid Handlers
- Mechanical Pipettes
- Electronic Pipettes
- Multi-Channel Pipettes
- Positive Displacement Pipettes & Dispensers
- Automated Pipetting
- Bottle-Top Dispensers
- Pipette Controllers
- Pipette Tips
- Automation Consumables
- Dispenser & Pipette Accessories
- Automation Accessories
- Dispenser & Pipette Services
Sorry, we couldn't find anything on our website containing your search term.

The Blank in Photometry
Lab Academy
- Molecular Biology
- Efficiency
- Photometry
- Quality
- Photometers
- Essay
Measuring the blank – and how to get it right
Determining blank, or zero, values is an important step in all photometric measurements. It serves the calibration of the photometer, which is thus set to “zero”. The blank value should encompass all those components which are included in the measurement of the sample, with the sole exception of the substance to be analyzed. Blank measurements are carried out in the same instrument, under the same experimental conditions, as the actual sample measurement. This approach will minimize any possible influence that the instrument, as well as the environment, may exert on the determination of the absorbance value of the sample. In addition, auto-absorbance of both the cuvette and the medium, in which the analyte is dissolved, is automatically subtracted from the result. While cuvettes are specially designed to be as transparent as possible for the light of photometric wavelengths, every material is subject to auto-absorbance that could potentially distort the result. The same is true for the solvent used.The following must be considered for materials as well as for the experimental approach:
Cuvettes:
Cuvettes should display the lowest possible auto-absorbance for the wavelengths utilized in order to avoid compromising the photometric measurement range. This may be checked and verified via determination of the blank value of the cuvette. To this end, the cuvette is filled with demineralized water and measured against a blank value which is obtained by measuring the empty cuvette shaft. When using glass cuvettes, special care should be taken to detect and avoid possible damage as well as contamination of the surfaces. In order to obtain the most accurate value possible, blank and sample should be measured using the same cuvette. It is also important to ensure that the orientation of the cuvette inside the cuvette shaft remains the same for each subsequent measurement. Markings on the sides of the cuvettes, which identify the transparent windows, allow correct and consistent orientation of the cuvette inside the cuvette shaft (figure 1).
Read more
Read less

Figure 1: Vis cuvettes with markings which indicate optical windows
Medium/reagents
The blank should comprise the solvent in which the analyte is solubilized and, if applicable, all other reagents that are added to the sample prior to measurement. If the medium itself shows high absorbance within the selected range of wavelengths, the photometric measurement range will be compromised. If in doubt, the absorbance value of the respective medium, including all applicable reagents, may be determined via photometric means using demineralized water as the blank value.
If no blank measurement is performed, or if the “wrong” blank medium is selected, the measured absorbance value of the sample may change. This phenomenon is illustrated in figure 2 by means of DNA absorbance spectra. Figure 2A shows a representative absorbance spectrum of pure DNA. In this case, the same Tris buffer which had been used to solubilize the DNA was employed as the blank solution. In contrast, the very different spectrum that was obtained when water was used as the blank solution is depicted in figure 2B. Marked differences were observed between the calculated purity quotients of A and B, respectively, which, in turn, are suggestive of sample contamination in the case of B. In addition, mistakes made during the preparation of the blank may lead to erroneous calculations of sample concentrations.
The blank should comprise the solvent in which the analyte is solubilized and, if applicable, all other reagents that are added to the sample prior to measurement. If the medium itself shows high absorbance within the selected range of wavelengths, the photometric measurement range will be compromised. If in doubt, the absorbance value of the respective medium, including all applicable reagents, may be determined via photometric means using demineralized water as the blank value.
If no blank measurement is performed, or if the “wrong” blank medium is selected, the measured absorbance value of the sample may change. This phenomenon is illustrated in figure 2 by means of DNA absorbance spectra. Figure 2A shows a representative absorbance spectrum of pure DNA. In this case, the same Tris buffer which had been used to solubilize the DNA was employed as the blank solution. In contrast, the very different spectrum that was obtained when water was used as the blank solution is depicted in figure 2B. Marked differences were observed between the calculated purity quotients of A and B, respectively, which, in turn, are suggestive of sample contamination in the case of B. In addition, mistakes made during the preparation of the blank may lead to erroneous calculations of sample concentrations.
Read more
Read less

Figure 2: Absorbance spectra of DNA solubilized in Tris buffer using different blank solutions: A) blank: Tris buffer; B) blank: water
Preparation:
The importance of careful preparation and performance of blank measurements should not be underestimated; errors which ensue during blank measurements may lead to incorrect measurements. Similar to the sample, the blank is subject to the minimum volume which is dictated by the choice of cuvette. Furthermore, the introduction of air bubbles into the cuvette (light path) must be avoided during the pipetting process, and the blank solution must be pure, i.e., it must be free from particulates and other contaminants. Negative absorbance values obtained for the sample, for example, may reveal problems with the blank solution or its measurement.
Measurement:
One blank solution can typically be used for multiple consecutive sample measurements. If, however, a considerable period of time is expected to elapse between individual measurements or if large numbers of samples are to be measured – for example, more than ten samples – a fresh blank measurement may be advisable. If measurement parameters change, a new blank measurement must be carried out in any case. Examples include differences in temperature or the type of cuvette used, as well as cases where the path length, the sample solution or the wavelengths are changed.
Analysis:
The software of modern (spectro)photometers automatically offsets the "measured value" of the blank against the absorbance values of the sample. In the case of older instruments, it may be necessary to manually subtract the blank value from the measured value of the sample.
The importance of careful preparation and performance of blank measurements should not be underestimated; errors which ensue during blank measurements may lead to incorrect measurements. Similar to the sample, the blank is subject to the minimum volume which is dictated by the choice of cuvette. Furthermore, the introduction of air bubbles into the cuvette (light path) must be avoided during the pipetting process, and the blank solution must be pure, i.e., it must be free from particulates and other contaminants. Negative absorbance values obtained for the sample, for example, may reveal problems with the blank solution or its measurement.
Measurement:
One blank solution can typically be used for multiple consecutive sample measurements. If, however, a considerable period of time is expected to elapse between individual measurements or if large numbers of samples are to be measured – for example, more than ten samples – a fresh blank measurement may be advisable. If measurement parameters change, a new blank measurement must be carried out in any case. Examples include differences in temperature or the type of cuvette used, as well as cases where the path length, the sample solution or the wavelengths are changed.
Analysis:
The software of modern (spectro)photometers automatically offsets the "measured value" of the blank against the absorbance values of the sample. In the case of older instruments, it may be necessary to manually subtract the blank value from the measured value of the sample.
Read more
Read less
Related links
Read more
Read less