Technical Data
Features
Technical Data
| Specifications | DASGIP® System for Cell Culture |
|---|
| Application | Cell culture |
| Number of parallel units | Up to 16 |
| Software | DASware control, other DASware optional |
| User Interface | Process computer with monitor |
| Typical power consumption (4-fold system without process computer) |
480 W (at 230 V)/ 432 W (at 115 V) (595 W/552 W without DASGIP Bioblock) |
| Dimensions (W × D × H, 4-fold system without process computer) | 140 x 70 x 53 cm (55 x 28 x 21 in) |
| Typical weight |
78 kg (60 kg without DASGIP Bioblock) |
| Bacteria/Yeast/Fungi | yes |
| Mammalian/Animal | yes |
| Insect | yes |
| Interchangeable Vessels | yes |
| Validatable | no |
| Stem cells | yes |
|
Bioreactor type |
|
| Vessels | Glass/single-use vessels |
| Sterilization |
Autoclavable/ pre-sterilized |
| Total volume | 1.5 L to 2.3 L |
|
Agitation |
|
| Drive | Direct/magnetic overhead drive |
| Speed ranges |
30 – 1,250 rpm (standard)/ 100 – 1,600 rpm (optional) |
| Impellers | Pitched-blade |
|
Gassing |
|
| Gas supply | TMFC; overlay and/or sparger |
| Standard gas flow rates | 0.1 – 50 sL/h, 0.1 – 40 sL/h CO2 (DASGIP MX4/4, other modules available) |
| Standard gas mixing | Air, N2, O2 and/or CO2 |
|
Feeding |
|
| Feed lines per vessel | Up to 8 |
| Standard feed rates (depending on tube diameter) |
0.3 – 9.5 mL/h to 13 – 420 mL/h (DASGIP MP8) 10 – 70 mL/h to 0.4 – 5 L/h (DASGIP MP4) |
|
Monitoring and Control |
|
| Temperature control | Heating and cooling integrated in DASGIP Bioblock (additional cooling fingers optional) |
| Standard temperature range (at 25° C room temp.) | 5 K above cooling agent temperature – 99°C (DASGIP Bioblock)1) |
| pH control | − |
| DO control | Cascade (O2 concentration, gas flow rate), and other set-ups |
| ORP (redox) measurement | - |
| Level/foam | Optional |
| OD measurement |
Optional (DASGIP OD4) |
| Exhaust condensation | Water-cooled or liquid-free (Peltier with DASGIP EGC4) |
| Exhaust analysis |
Optional (DASGIP GA4) |
|
1)Application-specific operating conditions apply. |
|