Software Downloads

Software and Firmware Updates

Software/Firmware revisions are made available at no charge through this web page so that all users can benefit from the latest performance enhancements.
The easy upgrade procedure takes only a few minutes to perform.

The software update tool for electronic pipettes and dispensers will be available shortly.

Updates for Existing CPUs

Analog Input/Output Module Updates

Date/File/Version Description of Changes
30-Jan-2019
Update_M1372_B.00.zip
Version Analog B.00
Analog I/O software revision B.00 includes the ability to recognize and control up to 8 Analog I/O modules from a single installed version, and addresses previous issues with communication loss.

BioCommand® Batch Control Updates

NOTE: This update is required for any current customer planning to use BioCommand witha BioFlo® 320. The enhancements made to are specific to the TCP/Modbus protocol used in the BioFlo® 320. There is no change to the functionality of BioCommand for all other New Brunswick™ bioprocess equipment.

NOTE: It is recommended that all customers update their software, however, customers running Revision C BioCommand software packages with all other New Brunswick™ bioprocess equipment are not required to install the update.

WARNING: Do not attempt to install this update on BioCommand® Batch Control Plus installations. Database files are not compatible and may result in loss of data. For BioCommand® Batch Control Plus updates, please contact your local sales/service representative.

Date/File/Version Description of Changes
30-Jan-2019
Update_BC_TT_BC_Version_D.zip
Version D
Revision D includes the following enhancements:
• Communication over TCP/Modbus protocol
• Addition of proportional and integral (P&I) values to programming section
• Addition of control mode selection to programming section
• Addition of P&I, deadband, and totalizer values to Batch Summary,    Datalog, and Reports

BioCommand® SFI Updates

Date/File/Version Description of Changes
22-May-2013
M1291-1252 
Rev C.00 - Update.zip

Version C
The updated version of BioCommand SFI allows for the correct interperation of values when used with European operating systems while continuing to report values on US operating systems. This improved version also allows BioCommand SFI to be used with up to 2 each New Brunswick S41i CO2 Incubator Shakers.
It trends and archieves critical data including temperature, speed and gas concentration but does not provide batch control.

BioCommand® Track and Trend Updates

NOTE: This update is required for any current customer planning to use BioCommand witha BioFlo® 320. The enhancements made to are specific to the TCP/Modbus protocol used in the BioFlo® 320. There is no change to the functionality of BioCommand for all other New Brunswick™ bioprocess equipment.

NOTE: It is recommended that all customers update their software, however, customers running Revision C BioCommand software packages with all other New Brunswick™ bioprocess equipment are not required to install the update.

WARNING: Do not attempt to install this update on BioCommand® Batch Control Plus installations. Database files are not compatible and may result in loss of data. For BioCommand® Batch Control Plus updates, please contact your local sales/service representative.

Date/File/Version Description of Changes
30-Jan-2019
Update_BC_TT_BC_Version_D.zip
Version D
Revision D includes the following enhancements:
• Communication over TCP/Modbus protocol
• Addition of proportional and integral (P&I) values to programming section
• Addition of control mode selection to programming section
• Addition of P&I, deadband, and totalizer values to Batch Summary, Datalog, and Reports

BioFlo®/CelliGen® 115 Updates

Date/File/Version Description of Changes

30-Jan-2019
Update_M1369_5.01.zip
Version 5.01 (E)

Software revision 05.01 incorporates the following solutions/additions to improve the performance of your bioprocess controller:

• Update for Corporate Identity
• Activation of Trending Module

BioFlo® 120 Updates

Date/File/Version Description of Changes

31-Jan-2019
Update_1390_
02.07.0002.zip

Version 02.07

Software revision 02.07 incorporates the following solutions/additions to improve the performance of your bioprocess controller: 

NOTE: This is a critical software update as IP addressing issues may lead to board failure, requiring board replacement

• Resolves Static IP addressing issues if network settings contains an IPv6 IP address or invalid Netmask.

BioFlo® 310, BioFlo 415 & CelliGen® 310 Updates

Date/File/Version Description of Changes
30-Jan-2019
Update_M1297
_12.04.zip

Version 12.04

Software revision 12.04 incorporates the following solutions / additions to improve the performance of your bioprocess controller:

  • Update for Corporate Identity
  • Includes updates from Revision L.02
  • Improved driver installation for RS-232 / USB device for externalscale interface (4 or 8 port)
  • Gas flow scaling has been added to all 3 or 4 gas configuration for CO2 gas
  • IP address function, allowing remote connectivity to RPC controller
 

BioFlo® 320 Updates

Date/File/Version Description of Changes
18-Nov-2020
Update_1379_02.08.0008.zip
 
Software revision 02.08 incorporates the following solutions/additions to improve the performance of your bioprocess controller: 
  • Issue found with manual network name affecting Static/DHCP setup. Network name detected from Netsh function
  • Delta-V Optical DO calibration issue resolved
  • Pump direction tag added to Modbus
  • 4-100RPM pumps change implemented
 

BioFlo®/CelliGen® 510 Updates

Date/File/Version Description of Changes

30-Jan-2019

UPDATE_M1361
_02.02.2006.zip

Version 02.02.0006

Software revision 02.02.0006 incorporates the following solutions/additions to improve the performance of your bioprocess system:
• TY-35B Overlay Value was always off (Close) on sterilization mode. This was resolved to allow TY-35B to open when the temperature falls below 121°C

BioFlo® 610 Updates

Date/File/Version Description of Changes

28-Oct-2019
BioFlo 610 update.zip

Software revision 02.05 incorporates the following solutions/additions to improve the performance of your bioprocess controller: 
  • Detection of touchscreen type for calibration purposes
  • Updated Modbus registers
 

CelliGen® BLU (1st Generation) Updates

NOTE: This update is specific to 1st generation CelliGen BLU cabinets. Please check the model of your CelliGen BLU cabinet. If your CelliGen BLU model number begins with the prefix "M1363", please follow the steps on this page to complete your update. If your CelliGen BLU model number begins with the prefix "M1374", please see the update CelliGen BLU (2nd Generation)

Date/File/Version Description of Changes
30-Jan-2019
Update_M1363_C.00.zip

CelliGen®BLU (first generation only) software revision C.00 provides the following resolutions and/or enhancements:

• Manual gas mixing total percent correction
• Improved RS-232 communication for scale driver
• Implementation of CO2 scaling for dedicated TMFC
• IP Network capability
• Improved storage of optical pH parameters

CelliGen® BLU (2nd Generation) Updates

NOTE: This update is specific to 2nd generation CelliGen BLU cabinets. Please check the model of your CelliGen BLU cabinet. If your CelliGen BLU model number begins with the prefix "M1374", please follow the steps on this page to complete your update. If your CelliGen BLU model number begins with the prefix "M1363", please see the update for CelliGen BLU (1st Generation).

Date/File/Version Description of Changes

30-Jan-2019
Update_M1374_2.02.zip
Version 2.02

Software revision 02.02 incorporates the following solutions / additions to improve the performance of your bioprocess controller:

Update for Corporate Identity

Update default Proportional and Integral values for pH loop

RackScan Analyze

NOTE: The RackScan Analyzer software is used to read 2D data matrix codes on consumables and to transfer the information contained in the 2D data matrix codes for further use in software applications.
The RackScan Analyzer software was developed to be used with RackScan b and RackScan s.

Date/File/Version Description of Changes
20-Dec-2019
RackScan Analyzer.zip
Version: 3.15 (build 3129)

 
The zip-file includes the following files:
  • Eppendorf RackScan Analyzer Installer
  • Software Manual including installation guide
 

RS-232 OPC Server Updates

Date/File/Version Description of Changes
30-Jan-2019
M1295 RS232 Device Interface Kit Rev B.00 - Update.zip
Version B
Update resolves communication errors with PC COM ports.

SciVario® Updates

Date/File/Version Description of Changes

24-Nov-2020

SciVario twin Software and Firmware Update 1.0.0.356.zip
Version 1.0.0.356

SciVario twin controller software and firmware revision 1.0.0.356 incorporates the following solutions/additions to improve the performance of your bioprocess controller:

  1. Implemented prohibition of SOM Bootloader downgrade for future releases
  2. Correction of agitation control parameters to improve performance
  3. Prevention of small pumps switching to maximal process value
 

ThermoMixer® C, ThermoMixer® F, and ThermoStat® C Updates

Date/File/Version Description of Changes
3.5.0-ThermoMixer_Autoupdate.zip
 

Software version 3.5.0 introduces a new application feature for thawing cells which is exclusively available for the Eppendorf ThermoMixer® C in combination with the new Eppendorf SmartBlock™ cryo thaw. Furthermore there are several minor additions, updates and fixes for the Eppendorf ThermoMixer® C, ThermoMixer® F, and ThermoStat® C such as

•    User notifications by LEDs and simple melodies in special programs (e.g. “Thawing cells”)
•    Mixing programs saved on program keys can be reset to their defaults (only ThermoMixer C)
•    Extended communication with VisioNize box
•    Translations for dialogs which were previously only available in English
•    Adapted lengths of display texts to better fit into GUI
•    Improved memory handling (see warning below)
•    Fixed scrolling behavior in menus with many entries