MENÚ
AR | ARS
AR | ARS
-
- Todas las pipetas, dispensadores y manipuladores de líquidos automatizados
- Pipetas mecánicas
- Pipetas electrónicas
- Pipetas multicanal
- Pipetas y dispensadores de desplazamiento positivo
- Dispensadores de botella
- Controladores de pipetas
- Accesorios para dispensadores y pipetas
- Pipeteo automatizado
- Consumibles para automatización
- Accesorios de automatización
- Servicios de manipulación de líquidos y pipetas
Sorry, we couldn't find anything on our website containing your search term.
Está a punto de abandonar este sitio.
Por favor, tenga en cuenta que su carro de la compra actual no ha sido guardado todavía y no podrá ser restablecido en el nuevo sitio ni cuando regrese. Si desea guardar su carro de la compra, inicie sesión en su cuenta.
Sorry, we couldn't find anything on our website containing your search term.
96-Wells at Once – Faster and More Reproducible Cell Assays
Academia de laboratorio
- Biología celular
- Reproducibilidad
- Automatización
- Eficiencia
- Estaciones de trabajo de manejo de líquidos
- Ensayo
Cell Based Assays are an established standard tool in life science and performed in 96 and 384-well plates. They require multichannel pipetting to ensure quick, efficient, and reproducible handling – especially when multiple plates need to be processed. A manual 12-channel pipette, an electronic 12-channel pipette and a semi-automated 96-channel electronic pipette (epMotion 96) were compared for cell seeding, cytotoxic agent and assay reagent addition.
Cell viability was evaluated after treatment with a staurosporine dose-response curve. Profiles generated were comparable for each instrument used for cell seeding and reagent adding. Nevertheless, with a manual pipetting system, the assay reproducibility was affected as indicated by higher standard deviations.
Leer más
S/B and Z´factor calculated for CellTubes - Blue Cell Viability Assay according to the instrument used for cell seeding and reagent adding
Intra- and inter-plate CVs calculated for CellTubes- Blue Cell Visability Assay for the respective instruments used
With all assay performance parameters studied (signal-to-background ratio, Z’-factor, intra-plate and inter-plate CV), it was demonstrated that the assay robustness and reproducibility were significantly better when an electronic pipette (epMotion 96 or electronic pipette) was used.
Leer más
Among all instruments tested, the epMotion 96 appears as the fastest option since a complete cell-based assay requires only 4 minutes, that means 5 times less hands-on time than what is needed with the manual pipette.
- to read the whole Application Note.
- to read the whole Application Note.
Leer más