MENU
BR | BRL
BR | BRL
-
- Todas as centrífugas
- Centrífugas de bancada
- Centrífugas de chão
- Centrífugas refrigeradas
- Microcentrífugas
- Centrífugas multifuncionais
- Centrífugas de alta velocidade
- Ultracentrífugas
- Concentrador
- Consumíveis de alta velocidade e de ultracentrifugação
- Tubos de centrifugação
- Placas de centrifugação
- Software de gestão de equipamentos
- Gestão de amostras e informações
-
- Todas as pipetas, dispensadores e equipamentos para manuseio de líquidos automatizados
- Pipetas mecânicas
- Pipetas eletrônicas
- Pipetas multicanal
- Pipetas de deslocamento positivo e dispensadores
- Ponteiras de pipeta
- Dispensadores de frascos
- Controladores de pipeta
- Acessórios para dispensadores e pipetas
- Pipetagem Automatizada
- Consumíveis de automação
- Acessórios de automação
- Serviços para pipetas e equipamentos para manuseio de líquidos
Sorry, we couldn't find anything on our website containing your search term.
Sorry, we couldn't find anything on our website containing your search term.
Calibration and Adjustment of Dispensing Systems in the Laboratory
Lab Academy
- Biotechnology
- Pipetting & Dispensing
- Dispensers
- Services
- Pipettes
- Essay
Calibrating pipettes regularly is important to secure a precise and accurate pipetting result. Executing a proper calibration according to ISO 8655 requires special ambient conditions like a temperature of 15-30 °C and air humidity of ~50 % in the laboratory.
The atmospheric pressure has to be considered for an exact calculation of the systematic and random error. Furthermore a calibrated analytical balance used for calibration in combination with an evaporation trap is essential. The evaporation trap is important when measuring small volumes < 10 µL. Each step of the calibration from pre-wetting the pipette tip, over pipetting angle to dispensing speed is critical and needs careful attention. This text guides you through the background of a good calibration and explains all steps to enable you to perform it on your own.
The figure shows the random and systematic measurement deviation for a 500 μl sample. Pipetting results can be imprecise, precise but inaccurate and ideally precise and accurate.
The figure shows the random and systematic measurement deviation for a 500 μl sample. Pipetting results can be imprecise, precise but inaccurate and ideally precise and accurate.
Read more
Read less
A good calibration can be compared to a target where the bull's eye should be hit and stands for a precise and accurate pipette. Each dot is a single pipetting event. The more distant each dot is from the bull's eye the less precise is the pipette. And the more scattered the dots are, the less accurate is the pipette.
Read more
Read less
Execution of a proper gravimetric pipette calibration using an analytical balance and an evaporation trap.
If you want to read more about calibration and adjustment of dispensing systems in the laboratory, please download our Userguide here!
If you want to read more about calibration and adjustment of dispensing systems in the laboratory, please download our Userguide here!