Measuring the Oxygen Transfer Rate (OTR) of Fermentation Bioreactors
Ying Yang, Marcus Deinstadt, and Ma Sha Academia de laboratorio
- Bioprocesamiento
- Cultivo de microorganismos
- Bioproceso
- Recipientes
- Ensayo
The protocol outlines the procedure for determining OTR of fermentation bioreactors using the sulfite depletion method.
Eppendorf, Inc. Enfield, CT USA
Abstract
This short protocol can serve as a reference for the bioprocess professionals who are interested in OTR measurement of the large collection of Eppendorf fermentation bioreactors.
A protocol for kLa measurement of cell culture bioreactors is also available.
Leer más
Leer menos
Table of contents
Leer más
Leer menos
Introduction: What is OTR and how to determine it
What is the oxygen transfer rate
The oxygen transfer rate (OTR) describes the rate of oxygen delivery from gas into liquid such as an agitated suspension culture in a bioreactor. The typical unit for OTR is mmol O2/(L x h), i.e., the amount of oxygen transferred into one liter of culture per hour. OTR is a very important parameter of fermentation bioreactors which affects dissolved oxygen control in bioreactors and which affects the selection, design, and scale-up of a bioprocess [1].The OTR of a bioreactor provided by most manufacturers represents the maximum oxygen transfer capability of this specific bioreactor under one VVM air sparging. There are many factors that affect the bioreactor’s OTR including the physical properties of the gas and liquid, operating conditions, and geometric characteristics of the bioreactor.
The mass transfer of oxygen is usually a limiting factor in aerobic fermentation bioprocesses. Under oxygen limitation, cell growth and product formation can be affected [2]. To assess their oxygen transfer capabilities, the Eppendorf applications team determined the OTR of many fermentation bioreactors.
Determining OTR using the sulfite depletion method
Here in this short protocol, we describe in detail how we measure the OTR of an Eppendorf fermentation bioreactor [3]. We use the industry standard sulfite depletion method for our measurement [1]. Maximum agitation and 1 VVM air flow provide oxygen transfer to the system. With the addition of sodium sulfite (Na2SO3), oxygen then rapidly reacts as the oxidizing agent to generate sodium sulfate (Na2SO4) in the presence of the copper catalyst.Leer más
Leer menos
The chemical reaction is shown here:

Leer más
Leer menos
Materials used in this protocol
Equipment
- An Eppendorf autoclavable glass bioreactor or a BioBLU® f Single-Use Bioreactor . In this study, we used a 3 L and a 5 L autoclavable glass bioreactor with stainless-steel dished bottom and direct drive for BioFlo 320.
- BioFlo 320 control station
- Mettler Toledo analog polarographic DO sensor (12 mm diameter with 220 mm insertion depth)
- Mettler Toledo analog polarographic DO sensor (12 mm diameter with 320 mm insertion depth)
- Mettler Toledo ISM polarographic DO sensor (12 mm diameter with 320 mm insertion depth)
- Mettler Toledo ISM optical DO sensor (12 mm diameter with 320 mm insertion depth)
Chemicals
- Sodium sulfite anhydrous (Na2SO3, Fisher Scientific® , S430-10)
- Copper(II) sulfate pentahydrate (CuSO4·5H2O) solution at 80 g/L concentration (J.T. Baker, 1843)
Other materials
- Weigh dish, lab balance, stopwatch, funnel, transfer pipettes, Easypet® 3 electronic pipette controller.
Leer más
Leer menos
Experimental procedures for OTR measurement
Bioreactor assembly
Assemble the glass bioreactor as per the user manual or take a ready-to-use BioBLU f Single-Use Bioreactor. Fill the bioreactor with deionized (DI) water to its maximum working volume.DO sensor calibration
Install a DO sensor in the bioreactor. Calibrate the sensor as described in Short Protocol 40 [4].OTR determination
We kept the temperature at 30 °C and sparged gas at 1 VVM with 100 % air. The experimental conditions are summarized in Table 1.Leer más
Leer menos

Leer más
Leer menos
Leer más
Leer menos

Results
An example of OTR measurement using the 3 L autoclavable glass vessel
We first assembled the 3 L BioFlo 320 fermentation bioreactor as per the user manual and filled the bioreactor with DI water to its maximum working volume which was 3.75 L in this case. Then we installed the polarized analog DO sensor in the bioreactor head plate and calibrated the sensor as follows. We set the temperature at 30 °C and agitation at 1200 rpm which is the maximum agitation rate of the bioreactor recommended by the user manual and sparged gas composed of 100 % nitrogen into the bioreactor at 3.75 SLPM (1 VVM). When DO reading stabilized after 10–30 minutes, zero (0 %) was set. Then we changed the gas composition to sparge 100 % air into the bioreactor at 3.75 SLPM, waited for another 10–30 minutes for the DO reading to stabilize to set the span (100 %). The DO sensor calibration was then completed.We kept the temperature at 30 °C, agitation at 1200 rpm, and gas sparging at 3.75 SLPM with 100 % air for the OTR measurement. At the beginning of the experiment, DO was stable at 100 %. We opened a Pg 13.5 port on the bioreactor head plate and transferred 7.5 mL of the 80 g/L copper sulfate stock solution into the bioreactor as the catalyst. We then weighed 41.25 g Na2SO3 powder and added it to the bioreactor through a funnel as quickly as possible to reach a final Na2SO3 concentration of 11 g/L. The ports were left open after chemical addition. Time was tracked by a stopwatch from the time point when DO reached 50 % on the downward trend to the time point when DO bounced back to 50 % on the way up as shown in Figure 2. We read the present values of DO manually on the control interface of BioFlo 320 or through the trend graph auto-collected by the unit, both at a data logging rate of every 5 seconds. The experiment was carried out twice to provide a mean OTR of this 3 L glass bioreactor. In this duplicated experiment, the elapsed time t we tracked for DO recovery was 427 s and 431 s, corresponding to 0.1186 h and 0.1197 h, respectively. The weight of Na2SO3 powder was 41.25 g. Based on this, using the equations shown above, we calculated the mean OTR to be 366.3 mmol O2/(L x h) for the 3 L autoclavable BioFlo 320 glass bioreactor (Table 2).
Leer más
Leer menos
DO sensor type comparison for OTR determination
For these experiments, we evaluated different types of DO sensors to compare their performances in OTR measurement. We first assembled a BioFlo 320 5 L fermentation bioreactor as per the user manual and filled it with DI water to its maximum working volume, which is 5.625 L. The sensors were polarized and calibrated together using the process stated previously in Short Protocol 40 [4]. Two experiments were designed for such comparisons.We first compared the analog DO sensor with the digital ISM sensor. We installed the two sensors in the same bioreactor for the experiment, removing the pH sensor to allow for the extra DO sensor in the process. This experiment was repeated three times so an average OTR can be calculated for each sensor. We then ran a second experiment comparing an analog DO sensor with the optical DO sensor. All experimental conditions were set the same as described below. The bioreactor temperature was set to 30 °C and gas sparging was set to 5.625 SLPM corresponding to 1 VVM with 100 % air for the OTR measurement. Agitation was set at the maximum of 1200 rpm. Pg 13.5 ports were opened on the headplate for chemical addition. Then 11.25 mL of the 80 g/L copper sulfate solution were added to the bioreactor, followed by the addition of 61.875 g Na2SO3. The ports were left open after chemical addition. Time was tracked by a stopwatch from the time point when DO reached 50 % on the downward trend to the time point when DO bounced back to 50 % on the way up as shown in Figure 2. The calculated OTRs of the 5 L glass bioreactor using these different DO sensors are shown in the Tables 3 and 4. The averages and standard deviations of these OTR measurements are further illustrated in Figure 4.
Leer más
Leer menos

Leer más
Leer menos
Discussion
In summary, this short protocol describes the method for measuring and calculating OTR of a fermentation bioreactor under industry standard OTR measurement conditions using the sulfite depletion method. A fermentation bioreactor with a high OTR can effectively support robust microbial growth and desired product yield. Therefore, this short protocol can serve as a reference for the bioprocess professionals who are interested in OTR measurement of Eppendorf’s large collection of fermentation bioreactors.
Leer más
Leer menos
Literature
[2] Hall SM. Rules of Thumb for Chemical Engineers (Sixth Edition). 2018
[3] Kelleher B, Nori C, Voll K, Siddiquee K, and Sha M. Evaluation of the BioFlo® 320 process capabilities. Eppendorf Poster. 2014
[4] Willard S. A Guide to Calibration on the BioFlo® 120 and BioFlo® 320: Dissolved Oxygen Sensors. Eppendorf Short Protocol No. 40 . 2017.
Leer más
Leer menos
Eppendorf SE reserves the right to modify its products and services at any time. This short protocol is subject to change without notice. Although prepared to ensure accuracy, Eppendorf SE assumes no liability for errors, or for any damages resulting from the application or use of this information. Viewing the short protocol alone cannot as such provide for or replace reading and respecting the current version of the operating manual.