MENÚ
AR | ARS
-
-
-
-
- Advanced Therapies Week (ATW) 2025
- Forum Labo 2025
- SLAS Europe 2025
- Bioprocessing Summit Europe 2025
- Medlab Middle East 2025
- SLAS International 2025
- Biologics World Nordics 2025
- ASIA LABEX: The Lab Show 2025
- BioProcess International Europe 2025
- ISEV 2025
- Future Labs Live 2025
- Cell 2025
- ASIA LABEX: The Lab Show 2025
- Stem Cell Community Day 2025
-
-
-
-
- Advanced Therapies Week (ATW) 2025
- Forum Labo 2025
- SLAS Europe 2025
- Bioprocessing Summit Europe 2025
- Medlab Middle East 2025
- SLAS International 2025
- Biologics World Nordics 2025
- ASIA LABEX: The Lab Show 2025
- BioProcess International Europe 2025
- ISEV 2025
- Future Labs Live 2025
- Cell 2025
- ASIA LABEX: The Lab Show 2025
- Stem Cell Community Day 2025
AR | ARS
-
- Pipeteo manual & dispensación
- Pipetas mecánicas
- Pipetas electrónicas
- Pipetas multicanal
- Pipetas de desplazamiento positivo y dispensadores
- Pipeteo automatizado
- Dispensadores de botella
- Controladores de pipeta
- Puntas de pipeta
- Consumibles de automatización
- Accesorios para dispensadores y pipetas
- Accesorios de automatización
- Servicios para dispensadores y pipetas
Está a punto de abandonar este sitio.
Por favor, tenga en cuenta que su carro de la compra actual no ha sido guardado todavía y no podrá ser restablecido en el nuevo sitio ni cuando regrese. Si desea guardar su carro de la compra, inicie sesión en su cuenta.
Sorry, we couldn't find anything on our website containing your search term.
_700.jpg)
There Is No Need to Lubricate Pipettes After Each Cleaning or Autoclaving
Academia de laboratorio
- Biotecnología
- Seguridad y bioseguridad
- Pipeteo y dispensación
- Pipetas
- Ensayo
It is not obvious when a pipette needs lubrication because the lubricant is inside the pipette and cannot be seen unless the device is disassembled. Pipettes don’t start to squeak. The only noticeable change might be an abrasive feeling or jerking when pushing the operation button which could be a first hint to control the lubricant on the pipette’s piston.
Leer más
Leer menos
Videos not loading, because cookies have been rejected. Change your

Re-lubrication is only necessary when cleaning agents such as soap, liquid detergent, or organic solvents have been used and the lubricant was actively removed. After calibration lubrication is not necessary, it does not affect the lubricant. Furthermore, it is important to only use a very small amount of lubricant, almost invisible to the human eye. This thin layer of lubricant should be applied with a fine cotton swab. Another important detail is the spot where to apply the lubricant. Most operating manuals explain where to lubricate the pipette. For our models (Eppendorf Research® plus, Eppendorf Reference® 2 and Eppendorf Xplorer®), when pipettes have a nominal volume greater than 20 µl, lubricate the cylinder in the lower part and for a nominal volume less than 20 µL, lubricate the small piston directly.
Leer más
Leer menos