There Is No Need to Lubricate Pipettes After Each Cleaning or Autoclaving
Lab Academy
- Pipettes
- Safety & Biosafety
- Pipetting & Dispensing
- Essay
It is not obvious when a pipette needs lubrication because the lubricant is inside the pipette and cannot be seen unless the device is disassembled. Pipettes don’t start to squeak. The only noticeable change might be an abrasive feeling or jerking when pushing the operation button which could be a first hint to control the lubricant on the pipette’s piston.
Read less
Videos not loading, because cookies have been rejected. Change your

Re-lubrication is only necessary when cleaning agents such as soap, liquid detergent, or organic solvents have been used and the lubricant was actively removed. After calibration lubrication is not necessary, it does not affect the lubricant. Furthermore, it is important to only use a very small amount of lubricant, almost invisible to the human eye. This thin layer of lubricant should be applied with a fine cotton swab. Another important detail is the spot where to apply the lubricant. Most operating manuals explain where to lubricate the pipette. For our models (Eppendorf Research® plus, Eppendorf Reference® 2 and Eppendorf Xplorer®), when pipettes have a nominal volume greater than 20 µl, lubricate the cylinder in the lower part and for a nominal volume less than 20 µL, lubricate the small piston directly.
Read less
_700.png)
Eppendorf Korea
- Legal Notice
- Legal Notice & Privacy Policy
- Trademarks & Patents
- Cookie Info
- For commercial customers only. All prices VAT exclusive.
- 상호: 에펜도르프코리아 주식회사 대표자: Christian Wolfgang Groeger 주소: 서울시 강남구 테헤란로 322 서관 6층 (역삼동, 한신인터밸리 24빌딩) 전화번호: 1577-4395 전자우편주소: cs@eppendorf.kr 사업자등록번호: 220-88-27824 통신판매업번호: 2023-서울강남-00526 판매 및 납품 일반약관
- Imprint
- © 2023 Eppendorf SE